Download User`s Manual - BME Shared Lab Resource
Transcript
Chapter 4: Using Your Freeze Dryer Moisture in Collector If there is liquid in the collector chamber and either a REFRIGERATION button or the VACUUM button is pressed, neither function will start and an alarm will be activated. When MENU is pressed, the display will show: MOISTURE IN COLLECTOR When the collector is drained, the alarm will self-cancel. This feature prevents the inadvertent start of the vacuum pump when liquid is in the collector and thereby can extend the life of the vacuum pump. A few drops of liquid may remain in the drain hose after draining. When the vacuum pump is started these drops may be sucked into the collector chamber. This is normal. Product Service: Domestic 1-800-522-7658, International 816-333-8811 27