4. Setting up the device
Menu level
Parameterisation level
Timeout, tIout:
The monitoring of the data transfer is parametrised in seconds, up to max. 100
seconds. By entering 000, no monitoring takes place. The timeout can be adjusted
vom the smallest to the highest digit with the up and down keys [▲] [▼] and needs to
be approved digit per digit with [P]. After the last digit, the display changes back into
menu level.
Back to menu group level, rEt:
With [P] the selection is confirmed and the device changes into menu group
level „– fct –“ .
4.3.5. Analog output parameters 1
Menu group level
Menu level
Menu level
Parameterisation level
Selection reference analog output, OutPt:
The analog output signal can refer to different functions, in detail this are the current
measuring value, Min-value, Max-value or totaliser-/sum-function. If HoLd is selected
the signal of the analog output will be hold and processed just after deactivation of
HOLD. With [P] the selection is confirmed and the device changes into menu level.