Download Gasman USA Manual Is..
Transcript
Safety Certification: Gasman is classified to US standards by Underwriter Laboratories Inc as suitable for use in Class 1 Groups A, B, C, D. Division I locations. Before use read the manual to ensure a full understanding of how to operate the unit and as to any limitations of use. • • S AS IFI CL • ED OX14 1DY UNITED KINGDOM ONLY AS TO INTRINSIC SAFETY FOR USE IN HAZARDOUS LOCATIONS UL CLASS I, DIV I, GROUPS A,B,C,D TEMPERATURE CODE T4 66Y6 FOR INTRINSIC SAFETY PRECAUTIONS SEE MANUAL WARNING: UNDERSTAND MANUAL BEFORE OPERATING ONLY CHANGE BATTERY IN A NON HAZARDOUS AREA ° ° EEx ia IIC T4 Ex ia IIC T4 -4F/-20 C Ta 149F/65 C -4F/-20 C Ta 149F/65 C BASEEFA04ATEX0384 IECEx BAS 05.0039 1180 II 1G • • • • • S AS IFI CL ED OX14 1DY UNITED KINGDOM ONLY AS TO INTRINSIC SAFETY FOR USE IN HAZARDOUS LOCATIONS UL CLASS I, DIV I, GROUPS A,B,C,D TEMPERATURE CODE T4 66Y6 FOR INTRINSIC SAFETY PRECAUTIONS SEE MANUAL WARNING: UNDERSTAND MANUAL BEFORE OPERATING ° ° ONLY CHARGE BATTERY IN A NON HAZARDOUS AREA EEx ia d IIC T4 Ex ia d IIC T4 -4F/-20 C Ta 149F/65 C -4F/-20 C Ta 149F/65 C BASEEFA04ATEX0383 IECEx BAS 05.0038 1180 II 2G Warning To avoid possible personal injury: • Substitution of components may impair intrinsic safety. • Do not use the detector if it is damaged or something is missing, contact Crowcon immediately. • Use only sensor modules specifically designed for your Gasman. (Please see the section on replacement parts and accessories.) • Periodically, “bump test” the sensor with gas to confirm its ability to respond to gas by exposing the detector to a targetted gas concentration that exceeds the high alarm setpoint. Manually verify that the audible and visual alarms are activated. It is recommended that the detector be calibrated before first-time use, and then at least once every 90 days and only by qualified personnel. For use in potentially explosive atmospheres where oxygen concentrations do not exceed 21%(v/v). Do not turn off the detector during a work shift. Turning off the detector resets the time weighted average gas exposure values to zero. Ensure that the sensor is not blocked. High off-scale %LEL readings may indicate an explosive concentration. Only charge the battery using a Crowcon supplied charger. Do not use any other charger. Other chargers may damage the detector making it unsafe for use in hazardous areas. Do not change or charge batteries in a hazardous location. Doing so may lead to fire or explosion. Flammable Gas Units Only • Extended exposure of the flammable sensor to certain concentrations of combustible gases and air may stress the sensor element, which can seriously affect its performance. If an alarm occurs due to a high concentration of combustible gases, recalibration should be performed, or if needed, the sensor replaced. • Do not test the combustible sensor’s response using a butane lighter. Doing so may damage the sensor. • Prior to each days usage, sensitivity must be tested on a known concentration of methane equivalent to 25-50% of full scale concentration (accuracy must be within 0±20% of actual). Accuracy may be corrected by recalibrating the instrument. • Any rapid up-scale reading followed by a declining or erratic reading may indicate a gas concentration beyond the upper scale limit, which may be hazardous.