Serial Communications Data Request Mode of Operation %1000<CR><LF> #1000=ddd<CR> Reply Min 0 Max 1 Unit Instrument Status %1001<CR><LF> #1001=ddd<CR> 0 4 Column Temperature %1005<CR><LF> #1005=ddddd<CR> 0 10000 Analysis Number %1008<CR><LF> #1008=ddddd<CR> 0 32767 N/a GC Concentrat. Peak 1 %4001<CR><LF> #4001=ddddd<CR> 0 32767 PPB*10 GC Concentrat. Peak 2 %4002<CR><LF> #4002=ddddd<CR> 0 32767 PPB*10 GC Concentrat. Peak 3 %4003<CR><LF> #4003=ddddd<CR> 0 32767 PPB*10 GC Concentrat. Peak 4 %4004<CR><LF> #4004=ddddd<CR> 0 32767 PPB*10 GC Concentrat. Peak 5 %4005<CR><LF> #4005=ddddd<CR> 0 32767 PPB*10 GC Concentrat. Peak 6 %4006<CR><LF> #4006=ddddd<CR> 0 32767 PPB*10 GC Alarm Level Peak 1 %4101<CR><LF> #4101=ddd<CR> 0 3 N/a GC Alarm Level Peak 2 %4102<CR><LF> #4102=ddd<CR> 0 3 N/a GC Alarm Level Peak 3 %4103<CR><LF> #4103=ddd<CR> 0 3 N/a GC Alarm Level Peak 4 %4104<CR><LF> #4104=ddd<CR> 0 3 N/a GC Alarm Level Peak 5 %4105<CR><LF> #4105=ddd<CR> 0 3 N/a GC Alarm Level Peak 6 %4106<CR><LF> #4106=ddd<CR> 0 3 N/a o C/10 Notes: - Mode of Operation : 000 – VOC, 001 – GC - Instrument Status: 000 – Stand By, 001 – Sample Analysis, 002 – Calibration Analysis, 003 – Integration, 004 – Pause (sequence operation) - Concentration unit is parts per billion * 10, (PPB*10). This means for a concentration of 234.123 ppb the reply data will be 00023, and for a concentration of 234.123 ppm the reply data will be 23412. - GC Alarm Level Peak N: 0 – No alarm, 1 – Caution, 2 – Warning, 3 – Alarm. - Unknown requests preceded by % are echoed back. - All other unknown requests illicit no response. 5-3