Download DVC-2010 User`s Manual

Transcript
Appendix
Specifications CST-5000 / CST-5000i
CV-2.1
Sheath
(08-5101)
Insertion Tube Diameter
(Nominal)
CV-1.5
Sheath
(08-5102)
Overall Endoscope Length (Distal Tip to
Top of Control Body)
Insertion Tube Working Length w/ Sheath
5.6 mm (16.8 Fr) with sheath
5.2 mm (15.6 Fr) with sheath
650 mm
370 mm
Field of View in Air/Water
110° (air) / 76˚ (water)
Direction of View
Forward
Depth of Field
3 - 50 mm
Environmental Effects on Optical
Performance
None
Angulation
215° Up/140° Down
Working Channel Diameter
Channel Diameter
CV-2.1
Sheath
(08-5101)
CV-1.5
Sheath
(08-5102)
Operating Environment
Temperature
Relative Humidity
Air Pressure
Storage Environment
Temperature
Relative Humidity
Air Pressure
Mode of Operation
Electrical Safety
Thermal Safety
Electromagnetic Compatibility
Degree of Protection Against Electrical
Shock
Degree of Protection Against Invasion of
Liquids
2.1 mm (6.3 Fr)
1.5 mm (4.5 Fr)
50° to 104° F (10° to 40° C)
30 to 85%
700 to 1060 hPa
14° to 140° F (-10° to +60° C)
0 to 95%
700 to 1060 hPa
Continuous
IEC 60601-1 & IEC 60601-2-18
IEC 60601-1 & IEC 60601-2-18
IEC 60601-1-2
Type BF
Fully Immersible (as per
Reprocessing Instructions)
Table A-1: Specifications
CST-5000/5000i Flexible Video Cystoscope/Hysteroscope User’s Manual
59