Download Model 2002 Multimeter User`s Manual
Transcript
IEEE-488 Reference Note: To disable all messages from entering the Error Queue, send the following command: :stat:que:enab () Program fragment PRINT #1, "output 16; :stat:que:enab (0:999); enab?" PRINT #1, "enter 16" ' Enable all messages and query ' Get response message from 2002 DISable <list> :STATus:QUEue:DISable <list> Parameters Disable messages for Error Queue <list> = (numlist) where; numlist is a specified list of messages that you wish to disable for the Error Queue. Query Description :DISable? Query list of disabled messages On power-up, all error messages are enabled and will go into the Error Queue as they occur. Status messages are not enabled and will not go into the queue. This command is used to specify which messages you want disabled. Disabled messages are prevented from going into the Error Queue. Messages are specified by numbers (see Table 2-4). See :QUEue:ENABle for examples to express a numlist. Program fragment PRINT #1, "output 16; :stat:que:dis (-140:-150); dis?" PRINT #1, "enter 16" ' Disable messages and query ' Get response message from 2002 :CLEar :STATus:QUEue:CLEar Description Program fragment 3-136 Clear Error Queue This action command is used to clear the Error Queue of messages. PRINT #1, "output 16; :stat:que:cle" ' Clear Error Queue