Operation
14. Remove the sample tray from the adapter of the InnuPure and move it back into the
priming station.
15. Close the elution vessels.
IMPORTANT! Load the sample tray evenly
For an extraction protocol both adapters in the sample tray must be filled in order to be balanced and to allow for optimum movement of the tray inside the device to the necessary
positions.
28

If more than one reagent strip are processed during the extraction protocol, insert
two adapters into the tray and distribute the strips evenly between them.

If only one sample is extracted, then place an adapter with an empty strip onto the
second adapter in the sample tray.

If you process the samples in a reagent plate, populate the second adapter with an
empty plate.