Chapter 4 Signal Connections In single-ended mode (RSE and NRSE), signals connected to ACH<0..15> are routed to the positive input of the PGIA. In differential mode, signals connected to ACH<0..7> are routed to the positive input of the PGIA, and signals connected to ACH<8..15> are routed to the negative input of the PGIA. Exceeding the differential and common-mode input ranges distorts your input signals. Exceeding the maximum input voltage rating can damage the device and the computer. National Instruments is not liable for any damages resulting from such signal connections. The maximum input voltage ratings are listed in the Protection column of Table 4-2. Caution In NRSE mode, the AISENSE signal is connected internally to the negative input of the PGIA when their corresponding channels are selected. In DIFF and RSE modes, AISENSE is left unconnected. AIGND is an analog input common signal that is routed directly to the ground tie point on the devices. You can use this signal for a general analog ground tie point to your device if necessary. The PGIA applies gain and common-mode voltage rejection and presents high input impedance to the analog input signals connected to your device. Signals are routed to the positive and negative inputs of the PGIA through input multiplexers on the device. The PGIA converts two input signals to a signal that is the difference between the two input signals multiplied by the gain setting of the amplifier. The amplifier output voltage is referenced to the ground for the device. Your device A/D converter (ADC) measures this output voltage when it performs A/D conversions. You must reference all signals to ground either at the source device or at the device. If you have a floating source, you should reference the signal to ground by using the RSE input mode or the DIFF input configuration with bias resistors, see the Differential Connections for Nonreferenced or Floating Signal Sources section in this chapter. If you have a grounded source, you should not reference the signal to AIGND. You can avoid this reference by using DIFF or NRSE input configurations. Analog Input Signal Connections The following sections discuss the use of single-ended and differential measurements and recommendations for measuring both floating and ground-referenced signal sources. © National Instruments Corporation 4-9 6034E/6035E User Manual