Appendices Appendix F: Pressure Correction Sensor Pressure is used in the context that this is the pressure inside the sampling chamber when performing the measurement, i.e. it is the operating pressure of the sensor. Gas Pressure is used in the context that this is the pressure at which the dewpoint is to be calculated. Pressure Correction is used in the context that the values displayed signify the moisture content at some pressure (we refer to this as the ‘Gas Pressure’) different from the pressure at the sensor. Note that PPM, LBS and G/M3 readings are by definition unaffected by pressure correction because only the pressure at the sensor affects their value. While °C and °F are affected by pressure correction by reporting what the dewpoint would be at the Gas Pressure when the dewpoint is what is measured at the pressure at the sensor. However, this also implies that whether pressure correction is applied or not the PPM, LBS and G/M3 readings are affected by the setting of the sensor pressure. The factory default settings are: 14.7psi for both sensor and gas pressure and pressure correction disabled. When Pressure correction is disabled all dewpoints are computed by assuming that both Sensor and Gas Pressures are 14.7psi. As all Xentaur dewpoint meters, the LPDT is equipped with the necessary software to perform pressure correction. However, for the lack of space, the LPDT does not have a Pressure Correct button. Therefore the user can not set or enable Pressure Correction, this must be done at the factory. Consult with your representative. page 39