Transcript
Specifications IF/NMES Combination Mode Specification IF/NMES Combination Preset Treatments Preset Treatment Total Treatment Time First 20 Minutes Second 20 Minutes IF Mode Sweep: 60 to 120 Hz Low Back 60 min. Ramp: 6 sec. up 6 sec. down IF Mode Sweep: 60 to 120 Hz Neck 60 min. Ramp: 6 sec. up 6 sec. down IF 89 3wave User’s Manual 926940_IF3Wave_Guts_Fr.indd 79 NM Mode 80 Hz 300 us Pulse Width Ramp: Up 2 sec. On 8 sec. Down 2 sec. Off 2 sec. NM Mode 80 Hz 300 us Pulse Width Ramp: Up 3 sec. On 6 sec. Down 3 sec. Off 2 sec. Last 20 Minutes IF Mode Sweep: 5 to 10 Hz Ramp: 12 sec. up 12 sec. down IF Mode Sweep:1 to 10 hz Ramp: 12 sec. up 12 sec. down 79 al IF 3WAVEE User’s Manua 2/27/14 8:57 PM