Transcript
User Manual – Rev D FCS Series II T1 0 Volt F1 F2 T2 Test no.: Standard: T1 (ms) F1 (Hz) F2 (Hz) T3 21 I 50 360 Fmax 22 II 50 Fmax 360 23 III 100 360 Fmax 24 IV 100 Fmax 360 25 V 200 360 Fmax 26 VI 200 Fmax 360 Fmax = 650 Hz for Group2/A(NF) Fmax = 800 Hz for Group3/A(WF) T2 = 20 msec T3 = 5 msec Figure 8-7: Power Interrupt for Group2/A(NF) and Group3/A(WF) VOLTAGE SURGE This test requires 160V output voltage. If the power source is set at the low voltage range, the high voltage range will be selected before the test starts. At the end of the test, the power source will be switched back to the low range automatically Seq. No. 1 2 3 4 5 RTCA 115 160 115 60 115 Group 1 115 160 115 70 115 Voltage Group 2 115 160 115 70 115 Group 3 115 170 115 70 115 Time ALL 5 Minute 30msec 5 Sec. 30msec 5 Sec. Table 8-5: Normal VoltageSurge Sequence The output voltage will follow the sequence in Table 8-5. The above sequence will repeat itself three times. Each repeat will start from sequence two. RTCA and Group 1 will run at 400 Hz. Group 2 and A(NF) will run at 360 Hz and 650 Hz. Group 3 and A(WF) will run at 360Hz and 800Hz. The frequency will return to the nominal setting when the test is completed. The ← key (backspace) will terminate the test at any time. California Instruments 120