5 Installation
Caution: Under filling may reduce the bath performance and may possibly
damage the bath.
5.4
Probe
Inspect the bath controller probe (see Figure 3 on page 24). It should not be
bent or damaged in any way. Reasonable caution should be used in handling
this probe as it contains a precision platinum sensor and is mechanically shock
sensitive. Dropping, striking, or other physical shock may cause a shift in resistance in the probe resulting in diminished bath accuracy. If damaged, the probe
can be replaced. Contact the factory for assistance.
Insert the probe into the 1 4 inch probe hole at the top left side of the bath lid.
The tip of the probe must be well immersed in the fluid. The probe connector is
plugged into the rear of the bath into the socket labeled “PROBE”.
5.5
Power
With the bath power switch off (see Figure 2 on page 23), plug the bath into an
AC mains outlet of the appropriate voltage, frequency, and current capacity.
See Section 3.1, Specifications, for power details.
Be sure the stirring motor power cord is plugged into the “STIRRER” socket at
the back of the bath (see Figure 4 on page 26).
6045/6050H
19