Download MP-0074, Rev E
Transcript
OPERATOR MANUAL.. Condition Duty Cycle Exceeded Alarm Visual Indicator Audible Signal ;Ò Double tone COOLING DOWN at the start and end of [count down] the cool down cycle Function / Resolution If the operator presses the trigger to activate the READY state before the duty cycle time, the device will display the warning message, and a countdown timer will be displayed showing the duty cycle time remaining. This will lock the device and the message will be displayed until the duty cycle is complete. The device will then return to the normal operation mode. Ensure the Handset is locked in place. Low Battery Alarm Incomplete Exposure Alarm System Failure Alarm Handset Failure RECHARGE HANDSET 5 audible tones Activates if trigger is released before the timed X-ray exposure finishes. INCOMPLETE A warning (PRESS ANY KEY) tone CALL MANUFACTURE HANDSET FAILURE This condition is cleared by pressing any button on the user interface panel or by tuning the power off then on. Depress the trigger for the duration of the exposure time. 0.38sec…….... SYSTEM FAILURE The RECHARGE HANDSET alert terminates after five seconds and goes into auto-shutdown. Replace the current Handset with a freshly charged Handset. NOTE: If battery voltage is lower than required for the X-ray exposure, the device will not allow the exposure. Descending tone Consult Section 6.2, Troubleshooting (6.2.8) or an Aribex authorized service center. 5 audible tones Consult Section 6.2, Troubleshooting (6.2.12) or an Aribex authorized service center. Alarm NOMAD Pro should not be operated if it has been dropped or performance degrades; it should be returned to Aribex for a safety check. 19