Transcript
Lake Shore Model 331 Temperature Controller User’s Manual Equipment Required for Calibration (Continued) Resistor Standards • Resistor standards with the following nominal values. If standards are not available, ¼ W 25ppm/oC metal film resistors can be used. They should have connectors attached to mate with two dual banana plugs for 4-lead measurement. • 0 Ω (short), 10 Ω, 100 Ω, 500 Ω, 1 kΩ, 5 kΩ, 100 kΩ. Miscellaneous • Dummy loads for warm up, 1 each for Diode/Resistor inputs: 6-pin DIN (240°) connectors – plug, with 100 kΩ resistors configured for 4-lead measurement. Calibration cable with 100 kΩ standard can be used. • Short length of uninsulated wire, 1 each for Thermocouple inputs. 8.9.2 Diode/Resistor Sensor Input Calibration NOTE: The thermocouple input calibration procedure in provided in Paragraph 8.9.3. Overview Each sensor input requires calibration. Sensor Inputs contain a current source which can supply 10 µA or 1 mA of current, only the 10 µA current sources are calibrated. They are calibrated by adjusting pots on the Model 331 Main Board. The sensor inputs contain multiple gain stages to accommodate the various sensors the Model 331 supports. The input circuitry is not adjusted during calibration. Instead, precision voltages and resistors are attached to each input and mathematical calibration constants are calculated and programmed into the Model 331. Constants are stored to compensate for both input offset and gain errors. Calibration Process 8.9.2.1 Sensor Input Calibration Setup and Serial Communication Verification Allow the Model 331 to warm up for at least 1 hour with 100 kΩ resistors attached to all inputs configured as diode/resistor and all thermocouple inputs shorted. Connect the Model 331 to the PC via the serial port. Verify serial communication by sending the *IDN? command and receiving the proper response from the Model 331. If the input not being calibrated is diode/resistor, leave a 100 kΩ resistor attached. If the other input is a thermocouple, leave a short across the input. 8.9.2.2 10 µA Current Source Calibration and 1 mA Current Source Verification Purpose To calibrate the 10 µA current source to be within the specified tolerance and verify operation of the 1 mA current source. Process 1. Configure the input for the Silicon Diode range. 2. Accurately determine the value of the 100 kΩ resistor using the DMM. Determine the calibration value by multiplying the actual resistance of the 100 kΩ resistor by 10 µA. Example: 100.050 kΩ × 10×10–6 A = 1.00050 V. 3. Attach the 100 kΩ resistor to the Model 331 input using proper 4-lead connection techniques, configure the DMM to read VDC and attach to the resistor. 4. Adjust the current source calibration pot, (R97 for Input A and R98 for Input B), on the Model 331 main board until the DMM reads exactly the value calculated in Step 2 to ±0.00002 VDC. 5. 1 mA Current Source Verification. Configure the input for the 100Ω Plat/250 range. 6. Accurately determine the value of the 1 kΩ resistor using the DMM. Determine the verification value by multiplying the actual resistance of the 1 kΩ resistor by 1 mA. 7. Attach the 1 kΩ resistor to the Model 331 input using proper 4-lead connection techniques, configure the DMM to read VDC and attach to the resistor. 8. Verify the voltage across to resistor to be within ±0.3% of the value calculated in Step 6. Service 8-11