Download XM Dynamic Measurement Module User Manual
Transcript
Configure XM Module in RSLogix 5000 Software Chapter 3 2. Configure the parameters, as necessary. In this field Values are Comments Enable Auto Trigger Check to enable Auto Trigger mode. The minimum signal amplitude for triggering is 2 volts peak-to-peak and the minimum frequency is 6 CPM (0.1 Hz). Using Auto Trigger mode may cause the tachometer to trigger on noise if the signal is very small. For example, you have 1 volt of noise on a 2 volt signal. To prevent this, make sure the % noise in the signal is less than the value entered in the Trigger Hysteresis. Clear to enable Manual Trigger mode. The value entered in Trigger Threshold is used as the trigger point. The minimum signal amplitude for triggering is 500 millivolts peak-to-peak and the minimum frequency is 1 CPM (0.016 Hz). Trigger Hysteresis Enter the amount of hysteresis around the trigger threshold. Enter a value between 0...50. In Auto Trigger mode, the value entered is a percentage of the peak-to-peak input signal. In Manual Trigger mode, the value entered is a voltage level. The hysteresis voltage is added to or subtracted from the threshold voltage to determine the hysteresis range. Trigger Level Enter the signal level to be used as the trigger value when in Manual Trigger mode. Trigger Slope Choose the input signal slope to be used with the trigger value. This parameter is dimmed in Auto Trigger mode. • Positive • Negative The trigger point of the tachometer defines 0° for phase measurement. If the tachometer is a square wave, the phase angles measured will vary by 180° depending on whether the Trigger Slope is set to positive or negative. DC High Limit Enter the maximum expected DC bias voltage from the transducer. DC Low Limit Enter the minimum, or most negative, expected DC voltage from the transducer. A voltage reading outside this range constitutes a transducer fault, which is indicated with the tachometer LED blinking red and the TachFault input tag. For information about the LEDs, see page 87. For information about the input tags, see page 91. Inhibit Zero Pulse Tachometer Check to enable Inhibit Zero Pulse Tachometer Fault. Fault Clear to disable Inhibit Zero Pulse Tachometer Fault. Controls whether a tachometer fault occurs if no pulses are detected on the tachometer signal. Fault Delay Enter a value between 1...64 seconds. Enter the number of seconds that the module should wait after the last valid pulse signal before it indicates a tachometer fault. Publication ICM-UM002F-EN-E - March 2013 69