Download caution - Thermo Scientific

Transcript
Section 3 Installation
Electrical
Requirements
DANGER
The circulator's construction provides protection against the risk
of electrical shock by grounding appropriate metal parts. The
protection will not function unless the power cord is connected to a
properly grounded outlet. It is the user's responsibility to assure a
proper ground connection is provided. 
The circulator is intended for use on a dedicated outlet. All circulators are
equipped with automatic thermally-triggered 20 Amp circuit protector.
The circuit protector is designed to protect the circulator.
Note If the circuit protector activates allow the circulator to cool before
resetting. Restart the circulator. Contact us if it activates again. 
CAUTION
The circulator's power cord is the disconnecting device, it must be
easily accessible at all times. 
Refer to the nameplate on the rear, upper-left-hand corner of the bath
for specific electrical requirements. Voltage deviations of ± 10% are
permissible. The outlet must be rated as suitable for the total power
consumption, see next page.
Thermo Scientific
3-3