Download Arbitrary Function Generator
Transcript
AFG-3000 Series User Manual Status Register Commands *CLS System Command Description The *CLS command clears all the event registers, the error queue and cancels an *OPC command. Syntax *CLS *ESE System Command Description The Standard Event Status Enable command determines which events in the Standard Event Status Event register can set the Event Summary Bit (ESB) of the Status Byte register. Any bit positions set to 1 enable the corresponding event. Any enabled events set bit 5 (ESB) of the Status Byte register. Note The *CLS command clears the event register, but not the enable register. Syntax Parameter *ESE <enable value> <enable value> Example *ESE 20 0~255 Sets a bit weight of 20 (bits 2 and 4). Query Syntax *ESE? Return Parameter Bit 0 200 Register Not used Bit 4 Register Message Available 1 Not used 5 Standard Event 2 Error Queue 6 Master Summary 3 Questionable Data 7 Not used