Download GasAlert Extreme
Transcript
GasAlert Extreme Calibration and Setting Alarm Setpoints If the span fails confirm that • gas is being applied to the sensor, • the sensor is detecting a sufficient gas concentration within 30 seconds, and • the gas concentration has not dropped significantly during the 2-minute span. If the span is still unsuccessful, attach a new gas cylinder. If the span continues to be unsuccessful, replace the sensor. Refer to Replacing the Battery or Sensor. Setting the Calibration Due Date After a successful calibration, the LCD displays the CAL. DUE screens and the number of days remaining before the next calibration is due. BW Technologies recommends that the detector be calibrated every 180 days (6 months). The detector is shipped with the factory default setting of 180 days. 8. Press D or E to scroll to the required value (1 to 365). 9. Press C to save the new value and automatically proceed to the TWA alarm setpoint screen. Note If the new value is not confirmed within 10 seconds after selecting, the detector automatically retains the existing value. The new value that is selected becomes the default value. Setting the TWA Alarm Setpoint Note To bypass and retain the current TWA alarm setpoint value, press C. The LCD automatically proceeds to the STEL alarm setpoint. Note To bypass the calibration due notification, press C. The LCD automatically proceeds to the TWA alarm setpoint. 45