R&S® ZNC Command Reference SCPI Command Reference Example: CALC4:PAR:SDEF 'Ch4Tr1', 'S11' Create channel 4 and a trace named Ch4Tr1 to measure the input reflection coefficient S11. CALC4:PAR:MEAS 'Ch4Tr1', 'a1' Change the measurement parameter of the trace and measure the wave quantity a1. CALC4:PAR:MEAS? 'Ch4Tr1' Query the measured quantity. The response is 'A1D1SAM'. Manual operation: See "S-Parameter" on page 113 CALCulate<Ch>:PARameter:SDEFine <TraceName>, <Result> Creates a trace and assigns a channel number, a name and a measurement parameter to it. The trace becomes the active trace in the channel but is not displayed. Note: To display the trace defined via CALCulate<Ch>:PARameter:SDEFine, create a diagram (DISPlay[:WINDow<Wnd>]:STATeON) and assign the trace to this diagram (DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED); see example below. CALCulate<Ch>:PARameter:MEASure changes the measurement result of an existing trace. To select an existing trace as the active trace, use CALCulate<Ch>:PARameter: SELect. You can open the trace manager to obtain an overview of all channels and traces, including the traces that are not displayed. Tip: This command has no query form. Use CALCulate<Ch>:PARameter:MEASure 'TraceName' to query the measurement result of the trace. CALCulate<Ch>: PARameter:CATalog? returns a list of all defined traces. Suffix: <Ch> . Channel number. <Ch> may be used to reference a previously defined channel. If <Ch> does not exist, it is generated with default channel settings. Setting parameters: <TraceName> Trace name, string variable, e.g. 'Trc4'. See "Rules for trace names" in "Trace Manager Table" on page 149. Trace names must be unique across all channels and diagrams. If a trace with the selected trace name already exists, the analyzer behaves as follows: If the existing trace is assigned to the same channel as the new trace, it is deleted. The new trace is not automatically assigned to a diagram area; see note above. If the existing trace is assigned to a different channel, no new trace can be created. The analyzer returns an error message. <Result> User Manual 1173.9557.02 ─ 13 Measurement result (string variable); see list of parameters below. 427
Related documents
R&S ZNC/ZND User Manual
ZVL Operating Manual
APPA`s New 60 Series Top of Low-Cost Digital Multimeters
Model 2002 Multimeter User`s Manual
Keithley 2000
User Manual
ZVB Quick Start Guide
The Pyrrho Book - The Pyrrho DBMS
USER MANUAL
Model 2316 RESISTOMAT
RC581 Configuration User Manual
Model 2000 Multimeter