Download Agilent 1290 Infinity II Variable Wavelength Detector User Manual

Transcript
Preparing the Module
Setting up an Analysis
Table 6
4
Choice of Priming Solvents for Different Purposes
Activity
Solvent
Comments
After an installation
Isopropanol
Best solvent to flush air out of the
system
When switching between reverse
phase and normal phase (both times)
Isopropanol
After an installation
Ethanol or Methanol
Alternative to Isopropanol (second
choice) if no Isopropanol is available
To clean the system when using
buffers
Bidistilled water
Best solvent to re-dissolve buffer
crystals
Best solvent to flush air out of the
system
Bidistilled water
After a solvent change
Best solvent to re-dissolve buffer
crystals
After the installation of normal phase
seals (P/N 0905-1420)
NOTE
Hexane + 5% Isopropanol
Good wetting properties
The pump should never be used for priming empty tubings (never let the pump run dry). Use
a syringe to draw enough solvent for completely filling the tubings to the pump inlet before
continuing to prime with the pump.
1 Open the purge valve of your pump (by turning it counterclockwise)
and set flow rate to 3 – 5 mL/min.
2 Flush all tubes with at least 30 mL of solvent.
3 Set flow to required value of your application and close the purge
valve.
NOTE
Pump for approximately 10minutes before starting your application.
Agilent 1290 Infinity II VWD User Manual
51