Download CY-7049
Transcript
t. uc od yo ur pr Total p53 ELISA Kit User’s Manual For Research Use Only, Not for use in diagnostic procedures ith Troubleshooting ew 1. The Human p53 Standard should be run in duplicate, using the protocol described in the Detailed Protocol. Incubation times or temperatures significantly different from those specified may give erroneous results. at ca m 2. Poor duplicates, accompanied by elevated values for wells containing no sample, indicate insufficient washing. If all instructions in the Detailed Protocol were followed accurately, such results indicate a need for washer maintenance. al th 3. Overall low signal may indicate that desiccation of the plate has occurred between the final wash and addition of Substrate Reagent. Do not allow the plate to dry out. Add Substrate Reagent immediately after wash. an u Reagent Stability eu se rm All of the reagents included in the CycLex Research Product CycLex Total p53 ELISA Kit have been tested for stability. Reagents should not be used beyond the stated expiration date. Upon receipt, kit reagents should be stored at 4°C, except the reconstituted p53 Standard must be stored at below -70°C. Coated assay plates should be stored in the original foil bag sealed by the zip lock and containing a desiccant pack. th Assay Characteristics re fe rt o 1. Sensitivity The limit of detection (defined as such a concentration of p53 giving absorbance higher than mean absorbance of blank* plus three standard deviations of the absorbance of blank: A blank + 3*SD blank) is better than 1.335 units/ml of sample. se * Dilution Buffer is pipetted into blank wells. nl y! Pl ea Eighty assays were evaluated and the minimum detectable dose (MDD) of p53 ranged from 0.926 – 1.637 units/mL. The mean MDD was 1.335 units /mL. The MDD was determined by adding three standard deviations to the mean optical density value of twenty zero standard replicates and calculating the corresponding concentration. eO 2. Specificity The antibodies in the p53 ELISA kit are highly specific of human p53, with no detectable cross-reactivity to mouse and rat p53. er en ce Pu rp os 3. Linearity To assess the linearity of the assay, samples containing and/or spiked with high concentrations of p53 were serially diluted with the Dilution Buffer to produce samples with values within the dynamic range of the assay. Fo rR ef Cat#: CY-7049 10 Version#: 140318