XL2 Remote Measurement
Reference Manual
Commands
Command Structure
The commands are send in ASCII format through the virtual COM port to your XL2 Audio and
Acoustic Analyzer. Every command transmission from your PC to your XL2 or vice versa must be
terminated with "CR LF" (Carriage Return, Line Feed).
The measurement commands are divided into six groups (i.e. "subsystems").
Subsystem
Function
*
Device status commands
INITiate
Status control for a measurement
MEASurement
Measurement result query commands
INPUt
Settings for Input signal path
CALIbrate
Microphone Calibration commands
SYSTem
System status commands
•
The XL2 accepts the short or any variant of the full form of the commands.
•
In the command list, the CAPITAL letters indicate the short form. However, the XL2 accepts both
lowercase and UPPERCASE letters, i.e. commands are not case-sensitive.
•
Multiple commands separated by semi-colons (";") are not supported.
•
Errors are stored in an error queue and can be queried with the “SYSTem:ERROr?” command.