Download NUFLO MC-III WP Flow Analyzer User Manual
Transcript
Appendix C MC-III™ WP Flow Analyzer Input Configuration Register (Decimal) Register (Hex) Data Type Access 2051 0803 Calibration Factor [9]* (pulses per configured factor unit) FP R/W 1.00 2053 0805 Calibration Factor [10]* (pulses per configured factor unit) FP R/W 1.00 2055 0807 Calibration Factor [11]* (pulses per configured factor unit) FP R/W 1.00 2057 0809 Calibration Factor [12]* (pulses per configured factor unit) FP R/W 1.00 2059 080B Standard Measurement Correction Factor (to be configured by software only) FP RO 1.00 2061 080D Low Flow Rate Cut-Off Threshold (in terms of configured flow rate units) FP R/W 1.00 Description Default *Calibration frequencies and calibration factors are associated in pairs—i.e., Calibration Frequency [1] and Calibration Factor [1]— and must be sorted in ascending order, based on calibration frequency values, prior to writing the values to the instrument. Therefore, Calibration Frequency [1] should be the smallest of all calibration frequencies. Output Configuration Register (Decimal) Register (Hex) 4000 0FA0 Pulse Output Source 0 = Test Mode/Disabled 1 = Enabled U16 R/W 0 4001 0FA1 Pulse Output Duration (time is register value multiplied by 10 ms) U16 R/W 10 (100 mS) 4002 0FA2 Pulse Output Decimal Point Location U16 R/W 2 4003 0FA3 Pulse Output Scale Factor (in terms of configured volume units) FP R/W 1.00 4005 0FA5 Analog Output Source 0 = Test Mode/Disabled 1 = Enabled U16 R/W 0 4006 0FA6 Analog Output Decimal Point Location U16 R/W 2 4007 0FA7 Analog Output Low Value (in terms of configured flow rate units) FP R/W 0.00 4009 0FA9 Analog Output High Value (in terms of configured flow rate units) FP R/W 1700.00 Important C-8 Description Data Type Access Default Two sets of 16-bit holding registers are provided (starting at 17001 and 8000). The “8000” set displayed in parentheses in the table on page C-9 should be used with master devices that require the holding registers to be in the range of 40000 to 49999.