6.10 QUERY Setup Screen – Custom Data Array Setup
The SETUP > QUERY screen allows the user to configure a custom digital data array for the
new Query Output, or for the European BH Bayern-Hessen protocol. The user can select
exactly which data parameters appear in the array, and in what exact order they appear. The
custom array can be as simple as a single concentration value, or it can be very
comprehensive including some parameters that are not even available in any of the standard
data files, such as the hourly span membrane check value. The setup of this custom array
does not affect any of the standard BAM data arrays. The query output is available in
firmware revisions 3.6.3 and later and requires the Report Processor. The BH protocol is
available in revision 5 series European firmware. A separate technical document for the
Bayern-Hessen protocol functions is also available. See Section 9.9 for instructions about
how to retrieve the Query output files.
N:08
01 TIME
04 AT
07 REF
10
13
16
SAVE
DATA QUERY FIELDS
02 CONC
03 FLOW
05 BP
06 RH
08 ERRORS
09
11
12
14
15
17
EXIT
The SETUP > QUERY Screen
The N: field sets how many values to include in the array. Up to 17 parameters can be
included. Increment the N value with the up/down arrow keys. For each increment of the N
value, another position in the array will be activated, starting with position 01.
Each position in the array can be changed to any desired parameter using the left/right keys
to select the position, and the up/down keys to scroll through the whole list of parameters. In
the example above, eight parameters are included and the shown parameters were selected
for each of the eight positions in the array. The available parameters are listed in the table
below:
Parameter
CONC
Q_STD
Q_ACT
STAB
REF
FLOW
CV
AT
BP
ANALOG 1
ANALOG 2
ANALOG 3
ANALOG 4
ANALOG 5
ANALOG 6
PM10s
Description
Concentration value for the last sample period.
Sample flow volume in cubic meters at standard conditions.
Sample flow volume in cubic meters at actual AT/BP conditions.
Stability measurement. Diagnostics use only. EU firmware only.
Reference span membrane mass measurement in mg/cm2.
Real time flow or average flow for the last sample.
Flow coefficient of variability for the last sample period. (Standard
deviation divided by the mean)
Average ambient temperature for the sample period.
Average barometric pressure for the sample period.
Average of analog met sensor channel 1 (user defined channel).
Average of analog met sensor channel 2 (user defined channel).
Average of analog met sensor channel 3 (user defined channel).
Average of analog met sensor channel 4 (usually filter RH).
Average of analog met sensor channel 5 (user defined channel).
Average of analog met sensor channel 6 (almost always AT).
PM10 concentration using standard conditions. Used in PM-coarse
systems where the coarse PM10 value is in actual conditions, but
BAM-1020-9800 REV K