Download Report 2 - CSA Group
Transcript
6.10 QUERY Setup Screen – Custom Data Array Setup The SETUP > QUERY screen allows the user to configure a custom digital data array for the new Query Output, or for the European BH Bayern-Hessen protocol. The user can select exactly which data parameters appear in the array, and in what exact order they appear. The custom array can be as simple as a single concentration value, or it can be very comprehensive including some parameters that are not even available in any of the standard data files, such as the hourly span membrane check value. The setup of this custom array does not affect any of the standard BAM data arrays. The query output is available in firmware revisions 3.6.3 and later and requires the Report Processor. The BH protocol is available in revision 5 series European firmware. A separate technical document for the Bayern-Hessen protocol functions is also available. See Section 9.9 for instructions about how to retrieve the Query output files. N:08 01 TIME 04 AT 07 REF 10 13 16 SAVE DATA QUERY FIELDS 02 CONC 03 FLOW 05 BP 06 RH 08 ERRORS 09 11 12 14 15 17 EXIT The SETUP > QUERY Screen The N: field sets how many values to include in the array. Up to 17 parameters can be included. Increment the N value with the up/down arrow keys. For each increment of the N value, another position in the array will be activated, starting with position 01. Each position in the array can be changed to any desired parameter using the left/right keys to select the position, and the up/down keys to scroll through the whole list of parameters. In the example above, eight parameters are included and the shown parameters were selected for each of the eight positions in the array. The available parameters are listed in the table below: Parameter CONC Q_STD Q_ACT STAB REF FLOW CV AT BP ANALOG 1 ANALOG 2 ANALOG 3 ANALOG 4 ANALOG 5 ANALOG 6 PM10s Description Concentration value for the last sample period. Sample flow volume in cubic meters at standard conditions. Sample flow volume in cubic meters at actual AT/BP conditions. Stability measurement. Diagnostics use only. EU firmware only. Reference span membrane mass measurement in mg/cm2. Real time flow or average flow for the last sample. Flow coefficient of variability for the last sample period. (Standard deviation divided by the mean) Average ambient temperature for the sample period. Average barometric pressure for the sample period. Average of analog met sensor channel 1 (user defined channel). Average of analog met sensor channel 2 (user defined channel). Average of analog met sensor channel 3 (user defined channel). Average of analog met sensor channel 4 (usually filter RH). Average of analog met sensor channel 5 (user defined channel). Average of analog met sensor channel 6 (almost always AT). PM10 concentration using standard conditions. Used in PM-coarse systems where the coarse PM10 value is in actual conditions, but BAM-1020-9800 REV K Page 55