Download DAQSTATION DX200 User`s Manual
Transcript
11.5 Setting the Alarm 3. Type Set the alarm type. [H]: Upper limit alarm [L]: Lower limit alarm [T]: Delay upper limit alarm [t]: Delay lower limit alarm Note If you select delay alarm (T or t) for the alarm type, you must set the alarm delay period. See section 6.3. 4. Value Set the value at which the alarm is activated. Pressing the [Input] soft key or one of the character/number input keys displays a window used to enter a numerical value. Enter a value and press the DISP/ ENTER key. For the procedures related to entering numerical values, see “Entering Numbers” on page 3-21. Relay On/Off Set whether or not to activate the relay output. If [On] is selected, output relay [Number] is displayed. Number Set the output relay number. For the correspondence between the output relay number and the output relay position, see section 2.4. 5.* 6.* * If the alarm output relay option (/AR1, /AR2, /A3, /A4, or /A5) is not installed, this setting is void. Confirming operation To confirm the new settings, press the DISP/ENTER key. To cancel the new settings, press the ESC key. A window appears for you to confirm the cancellation. Select [Yes] using the arrow keys and press the DISP/ENTER key. Explanation Alarm type Two types of alarms are available. 11-14 Name Symbol Description Upper limit alarm H An alarm occurs when the measured value becomes greater than or equal to the alarm value. Lower limit alarm L An alarm occurs when the measured value becomes smaller than or equal to the alarm value. Delay upper limit alarm T An alarm occurs when the measured value remains above or equal to the alarm value for the specified time period (delay period, see section 6.3, “Setting the Alarm Delay Period.”). Delay lower limit alarm t An alarm occurs when the measured value remains below or equal to the alarm value for the specified time period (delay period, see section 6.3, “Setting the Alarm Delay Period.”). IM 04L02A01-01E