Download EC9843 Operation Manual
Transcript
EC9843 NOy ANALYZER OPERATION MANUAL A built-in check verifies that the observed value is not different from the calibration value by more than 25%. If it is, no correction is made and an error message is sent to the EVENT LOG, setting the CALIBRATION ERROR flag. 8. The number displayed after NO SPAN RATIO and NO2 SPAN RATIO is the factor by which the instrument gains are multiplied to cause the display and output to agree with the span gas. You cannot set this number. This value is set to 1.000 any time the span is set manually via the front panel. (The assumption is that front panel adjustment is an instrument calibration, thus preventing compound adjustments.) 3.9.4 Description of the AZS Process The instrument will initiate a full zero/span cycle starting at the prescribed hour. The valve to admit zero air will be opened and the sample valve closed. The instrument will allow the cell to fill with the zero gas for 12 minutes. The display and outputs are updated with the actual instrument reading during the entire zero cycle. The zero value is for user reference only, and is never used by the analyzer to compensate readings. At the end of 12 minutes, the zero air valve is closed and the span gas valve is opened, admitting the span gas for 12 minutes. The display and outputs are updated with the actual instrument readings during the entire span cycle. The current value at the end of the span cycle is used to calculate the compensation value. If SPAN COMP is set to ENABLED, this is the value which is used to correct all subsequent readings to the calibration. The zero air valve is switched on for 3 minutes to purge the cell of span gas. At the end of 28 minutes, monitoring resumes including putting data in the average, etc. (The data averages are not updated during zero/span check.) 3.10 Calibration References 1. Quality Assurance Handbook for Air Pollution Measurement Systems. Volume 11-Ambient Air Specific Methods, EPA-600/4-77-027a, U.S. Environmental Protection Agency, Environmental Monitoring Systems Laboratory (MD-77), Research Triangle Park, NC 27711, 1977. 3-24 98437600 Rev D