Chapter 1
About the SCXI-1520
4.
If the cabled module is not listed in the correct slot, complete the
following troubleshooting steps:
a.
If the cabled module is not listed in the correct slot and the
slot is empty, click the drop-down listbox next to the correct
slot and select the cabled module. Configure the cabled
module following the steps listed in the SCXI Quick Start
Guide. Click OK.
b.
If another module appears where the cabled module should
be, click the drop-down listbox next to the correct slot and
select the cabled module. A message box appears asking you
to confirm the module replacement. Click OK. Configure the
cabled module following the steps listed in the SCXI Quick
Start Guide. Click OK.
•
Ensure that you have the highest priority SCXI module cabled to the
E/M Series DAQ device. Refer to the SCXI Quick Start Guide to find
out which SCXI module in the chassis should be cabled to the DAQ
device.
•
After checking the preceding items, return to the Troubleshooting the
Self-Test Verification section and retest the SCXI chassis.
If these measures do not successfully configure the SCXI system, contact
NI. Refer to the Technical Support Information document for contact
information.
Troubleshooting in Traditional NI-DAQ (Legacy)
SCXI-1520 User Manual
•
If you get the message Unable to test chassis at this time,
you have not designated at least one module as connected to a E Series
DAQ device. Refer to the Traditional NI-DAQ (Legacy) section of
Chapter 3, Configuring and Testing, and change the configuration of
the cabled module in the system from Connected to: None to
Connected to: Device x.
•
If you get the message Failed to find followed by the module
codes and the message Unable to communicate with chassis,
take the following troubleshooting actions:
–
Make sure the SCXI chassis is powered on.
–
Make sure the cable between the SCXI chassis and E Series DAQ
device is properly connected.
–
Inspect the cable connectors for bent pins.
–
Make sure you are using the correct NI cable assembly.
–
Test the DAQ device to verify it is working properly. Refer to the
DAQ device help file for more information.
1-8
ni.com