Download SAFETY TERMS AND SYMBOLS

Transcript
LCR-800 Series User Manual
LCR-800 Series User Manual
CONTENTS
PAGE
1.
INTRODUCTION……………………………………..……...
1
2.
PRECAUTIONS BEFORE OPERATION…….…………….
2-1.Unpacking the instrument………………….…………….
2-2.Checking the Line Voltage…………………..……………
2-3.Environment……………………………………..………..
2-4.Equipment Installation and Operation…………………
PANEL DESCRIPTION…...……………………..…………..
OPERATION…………….………………………………...….
4-1.Connects to DUT….………………..………….……..…...
4-2.Start-Up……….....……………………………..………….
4-3.Zeroing…………………………………………………….
4-4.Menu Functions.…..………………………………………
4-5.Measurement Conditions…..……………………………..
2
2
2
3
3
4
8
8
8
8
11
18
3.
4.
5.
SPECIFICATIONS…………………………………………... 35
6.
MESSAGE CODE.………………………………………….... 44
7.
MAINTENANCE....…………………………………………...
7-1.Cleaning……………………………………………………
7-2.Battery Replacement……………………………………...
7-3.Recall Calibration………………………………………...
45
45
45
46
8.
OPTION 1 (BIN Function for LCR-826/827/829 only)……..
8-1. BIN Functions for Components Sorting……………..….
8-2. BIN Setting Conditions.....……………………………….
8-3. BIN Range Setting: Component Sorting Range..............
48
48
50
56
9.
OPTION 2……………………………………………………..
9-1.On-line procedure…………………………………………
9-2.RS232 VIEWER Software Operation…………………...
9-3.The Configuration of Cable………………………………
65
65
69
74
i
SAFETY TERMS AND SYMBOLS
These terms may appear in this manual or on the product:
WARNING: Warning statements identify condition or practices
that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices
that could result in damage to this product or other property.
The following symbols may appear in this manual or on the product:
DANGER
ATTENTION
Protective
Frame or Chassis
High Voltage
refer to Manual
Conductor
Terminal
Terminal
ii
LCR-800 Series User Manual
LCR-800 Series User Manual
This cable/appliance should be protected by a suitably rated and approved
FOR UNITED KINGDOM ONLY
HBC mains fuse: refer to the rating information on the equipment and/or
user instructions for details. As a guide, cable of 0.75mm2 should be
NOTE: This lead / appliance must only be wired by competent persons
protected by a 3A or 5A fuse. Larger conductors would normally require
13A types, depending on the connection method used.
WARNING: THIS APPLIANCE MUST BE EARTHED
Any moulded mains connector that requires removal /replacement must be
destroyed by removal of any fuse & fuse carrier and disposed of
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
immediately, as a plug with bared wires is hazardous if a engaged in live
socket. Any re-wiring must be carried out in accordance with the
information detailed on this label.
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
As the colours of the wires in main leads may not correspond with the
colours marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with the letter E or by the earth symbol
or coloured
Green or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is
marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal
marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact
the supplier.
iii
iv
LCR-800 Series User Manual
LCR-800 Series User Manual
EC Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
1. INTRODUCTION
The precise LCR meter series are automatic, user programmable instruments
No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan
that provide high reliability and great precision for measuring a wide variety of
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
impedance parameters. The frequency range of LCR-819/829 covers from 12Hz
No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declare that the below mentioned products:
LCR-817/819/827/829/816/826/821
are herewith confirmed to comply with the requirements set out in the Council
Directive on the Approximation of the Law of Member States relating to
Electromagnetic Compatibility (89/336/EEC, 92/31/EEC, 93/68/EEC) and Low
Voltage Equipment Directive (73/23/EEC, 93/68/EEC).
For the evaluation regarding the Electromagnetic Compatibility and Low Voltage
Equipment Directive, the following standards were applied:
to 100kHz, LCR-817/827 from 12Hz to 10kHz, LCR-816/826 from 100Hz to
2kHz, and LCR-821 from 12Hz to 200kHz. The basic accuracy is 0.1% for LCR827/829/816/826, 0.05% for LCR-817/819/821. The measured results can be
displayed on the high quality LCD monitor with decimal points and units. The
measured result resolution is five full digits for Inductance (L), Capacitance (C),
Resistance (R) and absolute value of impedance (|Z|)
[four full digits for
Dissipation factor (D), Quality factor (Q), R with C, R with L, 2 digits after
EN 61326-1: Electrical equipment for measurement, control and laboratory use ––
EMC requirements (1997+A1: 1998)
Conducted and Radiated Emissions
Electrostatic Discharge
EN 55011 : 1998 Group I Class A
IEC 61000-4-2: 1995
Current Harmonic
Radiated Immunity
EN 61000-3-2: 2000
IEC 61000-4-3: 1995
Voltage Fluctuation
Electrical Fast Transients
EN 61000-3-3: 1995
IEC 61000-4-4: 1995
Surge Immunity
————————————————
IEC 61000-4-5: 1995
Conducted Susceptibility
————————————————
IEC 61000-4-6: 1996
Voltage Dips/ Interrupts
————————————————
IEC 61000-4-11: 1994
decimal dot for phase angle (θ)]. In the meantime, the LCD monitor is also shown
the control status and parameters of settings. The keypads are easy for menu
programming. The LCR Meters’ test fixture (option) is a method of convenient,
reliable, guarded 4-terminal/2 wires connection of radial and axial leaded
components to the LCR Meters. Test conditions can be stored and recalled from
internal memory that will reduce the setup time for measurement preparing.
Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC
Safety Requirements
IEC/EN 61010-1:2001
v
1
LCR-800 Series User Manual
LCR-800 Series User Manual
2. PRECAUTIONS BEFORE OPERATION
The normal ambient temperature range of the LCR Meters is from 10° to
50°C. To operate the instrument over this specific temperature range may
2-1. Unpacking the instrument
The product has been fully inspected and tested before shipping from the
factory. Upon receiving the instrument, please unpack and inspect it to
check if there are any damages caused during transportation. If any damage
is found, notify the bearer and/or the dealer immediately.
cause damage to the circuits.
Do not use the LCR Meters in a place where strong magnetic or electric
field exists as it may disturb the measurement.
2-4. Equipment Installation and Operation
Ensure there is proper ventilation for the vents in the LCR meters case. If
2-2. Checking the Line Voltage
The LCR Meters can be operated with AC power source between 100V
rated voltage and 240V rated voltage at a frequency of 50 to 60Hz, no AC
voltage selector is necessary. Power connection to rear panel is through an
AC inlet module comprised of an AC connector and fuse holder. To change
the fuse proceed as follows:
z
2-3. Environment
Remove the fuse holder by inserting a small flat head screwdriver
this equipment is used not according to the specification, the protection
provided by the equipment may be impaired.
WARNING: This is a Class A product. In a domestic environment
this product may cause radio interference in which case the user
may be required to take adequate measures.
behind the small tab to force the holder outward.
z
Install the correct fuse (slow-blow, 3A, 250Vac).
z
Re-install the fuse holder back into the LCR Meters AC inlet module,
push in and lock.
WARNING: To avoid electrical shock the power cord protective
grounding conductor must be connected to ground.
WARNING: To avoid personal injury, disconnect the power cord
before removing the fuse holder.
2
3
LCR-800 Series User Manual
LCR-800 Series User Manual
3. PANEL DESCRIPTION
(10). Compound key
z For
z Selects
(1). Power Switch
of LCD monitor will display the “INT.B” message. If the external DC
(2). Function key 一 F1
bias is selected, the bottom of LCD monitor will display the “EXT.B”
Soft key functions as indicated on the adjacent LCD monitor.
(3). Function key 一 F2
Soft key functions as indicated on the adjacent LCD monitor.
Soft key functions as indicated on the adjacent LCD monitor.
Enters menu display mode or exits sub menu back to main menu.
z For
making numerical entries as “-“ (the negative sign).
making numerical entries as labeled.
the “RANGE HOLD” mode on or off.
(8). Compound key
the “CONSTANT VOLTAGE” mode on or off.
(9). Compound key
making numerical entries as labeled.
the unit of Dissipation and Quality Factor in PPM.
(Note: The models of LCR-816 and 826 do not have this function.)
4
measurement sequence. Normally used in the “MANU” (Triggered)
mode.
z Selects
“AUTO” or “MANU” mode by pressing this key for 3 seconds at
least.
z The
making numerical entries as labeled.
the “TEST FREQUENCY”.
(13). START (Compound key)
z Starts
(7). Compound key
z Measures
the INTERNAL BIAS mode or EXTERNAL BIAS mode “ON” or
“OFF”.
z Inputs
(6). MENU key
z For
making numerical entries as labeled.
(12). Compound key
(5). Function key 一 F4
z Turns
z For
z Turns
(4). Function key 一 F3
z For
message.)
(11). Compound key
Soft key functions as indicated on the adjacent LCD monitor.
z Turns
the “INTERNAL BIAS” mode or “EXTERNAL BIAS” mode. (If
this key function is switched to “INTERNAL BIAS” mode, the bottom
Turns AC Power on or off.
z For
making numerical entries as labeled.
LCR Meters will process the measurement automatically, if the
“AUTO” mode is selected.
(14). ↵ key (ENTER)
This key enables programming of all special functions, test frequency, test
voltage, averaging, delay, and nominal value etc.
(15). ‧ Symbol key
Inputs the decimal point
5
LCR-800 Series User Manual
LCR-800 Series User Manual
(16). Numeral key—“2”
FRONT PANEL
(17). Numeral key—“3”
(18). Numeral key—“5”
(19). Numeral key—“6”
(20). Numeral key—“9”
(21). Primary Display
This line can display the measured Inductance, Capacitance, or Resistance
(LCR-821 has Ζ).
(22). Secondary Display
This line can display the measured Quality Factor or Dissipation or ESR or
EPR (LCR-821 has Θ).
(23). Instrument status or indicates measurement results based on entered test
limits.
(24). Test conditions
(25). Input terminals
BNC connectors, connects to device under test (DUT).
Connectors of the LCR Meters
BIAS
Lforce (current, low)
+
Lsense (potential low)
+
Hsense (potential high)
-
Hforce (current, high)
-
6
7
LCR-800 Series User Manual
LCR-800 Series User Manual
z
4. OPERATION
After the BAR at the bottom of LCD monitor is filled to the full, the zeroing
process is done.
z
4-1. Connects to DUT
If the zeroing process is successful, a message of “OK” will appear on the
LCD monitor. If failed, a message of “FAIL” will appear on the LCD
The LCR Meters utilize the structure of four wires measurement which allows
monitor.
accurate, easy, and stable measurements and avoids mutual inductance and
interference from measurement signals, noise and other factors inherent with
Short Circuit
other types of connections. For the accuracy of measurement, GOODWILL
z
The test cable should be connected or test fixture shorted (using a clean
copper wire, as short as possible).
produces the cable set and test fixture (option) for connection directly to the
front panel BNC connectors.
z
Press MENU key.
4-2. Start-Up
z
Press F1 key to select “OFFSET” menu.
z
Press F2 key to select short circuit zeroing (the “R/L OFFSET” is indicated
Connects the power cord of the LCR Meters to the mains socket-outlet. Presses
the POWER button of front panel to apply the AC power to the LCR Meters.
on the adjacent LCD monitor.).
z
4-3. Zeroing
After the BAR at the bottom of LCD monitor is filled fully, the zeroing
process is done.
In order to eliminate strayed capacitance and impedance of test cable during the
z
If the zeroing process is successful, a message of “OK” will appear on the
measurement, the LCR Meters should be zeroed to correct for test cable and/or
LCD monitor. If failed, a message of “FAIL” will appear on the LCD
test fixture errors before taking measurements. The corrections are calculated
monitor.
and stored in memory of the LCR Meters during the zeroing process. Open and
short circuit zeroing should be done for test cable and/or test fixture. For the best
accuracy, the test cable and/or test fixture should be zeroed once per day at least
and each time test cable or test fixture is changed in general.
The zeroing process of open and short circuits are following:
Test Condition:
Test voltage=According to the actual test voltage.
Test speed = According to the actual test speed.
R.H = OFF
C.V = According to the actual test status.
Open Circuit
z
The test cable or test fixture should be open with no component connected.
For the summary, the zeroing menu can be chosen through menu selection as
z
Press MENU key.
shown in Figure 4-1 above.
z
Press F1 key to select “OFFSET” menu.
z
Press F1 key to select open circuit zeroing (the “CAP OFFSET” is indicated
on the adjacent LCD monitor.).
8
NOTE: The “Open Circuit” and “Short Circuit” have to pass the test,
otherwise, the accuracy of the LCR Meters will become worse.
9
LCR-800 Series User Manual
L 1.2345H
Q 0.6789
TESTING
F : 1.000 kHz R.H OFF
C.V OFF
V : 1.000 V
AUTO MANU INT.B OFF
CAP. R/L OFFSET
SET SORT
LCR-800 Series User Manual
4-4. Menu Functions
SPEED
SLOW
All the LCR Meters’ programmable functions are controlled by the easy to use
DISPLAY
VALUE
menu displays. User can enter the menu mode by selecting the MENU key that
calls up four top level menus, OFFSET, SORT, SETTING and CALBRAT.
MODE
L/Q
CIRCUIT
SERIES
Each one of these is comprised of a sub menu list whose functions are described
Press menu
key
MENU
OFFSET
Press F1 key
to the zeroing
menu
in detail below. User can enter one of four functions by pressing the
corresponding function key (just adjacent LCD monitor, see figure 4-2).
CAP. R/L OFFSET
OFFSET
SET SORT
SORT
SORT
SET PARAMETER
SETTING
CALIBRATION
CALBRAT
SET PARAMETER
SETTING
CALIBRATION
CALBRAT
EXIT
EXIT
CAP
OFFSET
SHORT TEST
R/L
OFFSET
Press F2 key
to the bin sort
menu
Press F3 key
to the utility
menu
Press F4 key
to the
calibration
menu
Press MENU key
to exit menu screen
Press F1 KEY for
open test
OPEN TEST
Press F1 key
to the zeroing
menu
Figure 4-2: four main menu screen
Press F2 KEY for
short test
4-4-1. Primary & Secondary Display
For the LCR Meters, four combinations of two parameters can be measured and
displayed simultaneously.
One referred to the “Primary Display” (displayed
first) and the other to the “Secondary Display”. Depending on the component
EXIT
type the primary and secondary display could be L & Q, C & D, C & R or R and
Q, and add L & R and Z & θ to LCR-821.
Figure 4-1: Summary of zeroing menu
10
pressing F3 key as shown in Figure 4-3.
11
The parameter can be chosen by
LCR-800 Series User Manual
LCR-800 Series User Manual
Primary
Display
z
Secondary
Display
L 1.2345H
Q 0.6789
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
TESTING
F : 1.000 kHz R.H OFF
V : 1.000 V
C.V OFF
AUTO MANU INT.B OFF
CIRCUIT
SERIES
MENU
Press F3 key
successively in
order to select the
different
measurement mode
(R/Q, C/D, C/R,
L/Q, Z/θ, L/R)
Measurement Function
Measurement Mode
Series Equivalent Circuit
R/Q
C/D
C/R
Z / θ (degree)
L/R
9
9
9
9
9
Parallel Equivalent
Circuit
9
9
9
9
Suggested Test Conditions:
Inductors less than 10µH: Series, 100kHz.
Inductors from 10 µH to 1mH: Series, 10kHz.
Inductors from 1mH to 1H: Series, 1kHz.
Inductors greater than 1H: Series, 0.1kHz.
Figure 4-3. Primary & Secondary display
Capacitors less than 10pF: Parallel, 100kHz.
User can select R/Q for resistor measurement; select L/Q or L/R for inductor
Capacitors from 10 to 400pF: Series or Parallel, 10kHz.
measurement; select Z/θ for impedance measurement, select either C/D or C/R
Capacitors from 400 to 1µF: Series, 1kHz.
for capacitor measurement.
Capacitors greater than 1µF: Series, 0.1 or 0.12kHz.
Note: Only LCR-821 can select Z/θ and L/R measurement modes.
Resistor less than 1kΩ: Series, 1kHz.
4-4-2. Series & Parallel Equivalent Circuit
Impedance that is neither a pure resistance nor a pure reactance can be
represented at any specific frequency by either a series or a parallel combination
of resistance and reactance. Such representation is called “equivalent circuit”.
The component value of the “Primary Display” depends on which equivalent
circuit (series or parallel) is chosen. In normal, the component manufacturer
shall specify how a component is to be measured (usually series) and at what
frequency. Please refer to Figure 4-4.
Resistor from 1kΩ to 10MΩ: Parallel, 0.25kHz.
Resistor greater than 10 MΩ: Parallel, 0.03kHz
Unless for special reason, always select “Series” for capacitors and inductors.
This has traditionally been standard practice. For very small capacitance or
inductance, select a higher test frequency for better accuracy. For very large
capacitance or inductance, select a lower test frequency for better accuracy. For
dc resistance, select a lower test frequency to minimize ac effects.
Because the reactive component most likely to be represented in a low resistance
resistor is series inductance, the “Series” is selected for a resistor below about
1kΩ. If a resistor large than 10MΩ, select “Parallel” that because the reactive
component most likely to be represent in a high resistance resistor is shunt
capacitance. If the Q is less than 0.1, the measured Rp is probably very close to
the dc resistance.
12
13
LCR-800 Series User Manual
LCR-800 Series User Manual
The total loss of a capacitor can be expressed in several ways, including D and
“ESR” (Equivalent Series Resistance). “ESR” is typically much larger than
actual “ohmic” series resistance of the wire leads and foils that are in series with
the heart of a capacitor physically, because ESR includes also the effect of
dielectric loss. ESR is related to D by the formula: ESR = Rs = D/ωCs. Where ω
4-4-3. Measurement Displays
The measured results of the LCR Meters can be shown on the LCD monitor in
three ways: VALUE, DELTA%, or DELTA. User can press F2 key to select the
appropriate item for measurement. Please refer to Figure 4-5.
represents “omega” = 2 pi time frequency.
VALUE
Although it is traditional to measure series inductance of inductors, there are
The LCD monitor will display the measured value of both the primary and
situations in which the parallel equivalent circuit better represents the physical
secondary parameter, shown with decimal and units. The resolution of primary
component. For small “air-core” inductors, the significant loss mechanism is
display (L, C, R or Z) is five digits. The resolution of secondary display (D, Q, R
usually “ohmic” or “copper loss” in the wire, therefore the series circuit is
with C or R with L) is four digits. The resolution of secondary display (θ) is at 2
appropriate. Nevertheless, for an “iron core”, the significant loss mechanism can
digits after decimal point. The message “TESTING” is displayed when a test is
be “core loss”, therefore, the parallel equivalent circuit is appropriate which
in process.
being a better model of the inductor.
DELTA%
The “DELTA%” shows the percent deviation of the measured L, C, R or Z value
L 1.2345H
Q 0.6789
DELTA
The LCR difference is similar to the DELTA% except that the deviation is shown
DISPLAY
VALUE
in suitable units (ohms, henries, etc).
MODE
L/Q
TESTING
F : 1.000 kHz R.H OFF
C.V OFF
V : 1.000 V
AUTO MANU INT.B OFF
from a stored NOMINAL VALUE. The sign of deviation is indicated.
SPEED
SLOW
CIRCUIT
SERIES
Press F4 key to
select the Series
or Parallel circuit
MENU
L 1.2345H
Q 0.6789
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
Figure 4-4. Selections of Series & Parallel Circuit
TESTING
F : 1.000 kHz R.H OFF
C.V OFF
V : 1.000 V
AUTO MANU INT.B OFF
Press F2 key to select the
three different measuring
display ("VALUE",
"DELTA%", or "DELTA").
CIRCUIT
SERIES
MENU
Figure 4-5. Types of measurement display
14
15
LCR-800 Series User Manual
LCR-800 Series User Manual
4-4-4. Nominal Value
Allows entry of a “Nominal Value” for the primary parameter which is the basis
for the measurement result in “DELTA” or “DELTA %”. Accepts numerical entry
up to five digits with decimal. Units are depended on which measurement
displays selected.
Steps of “Nominal Value” input (Figure 4-6):
z
Press MENU key.
z
Press F2 key to select “SORT” menu.
z
Press F1 key to select “Nominal Value” (the “NOM.VAL” is indicated on
the adjacent LCD monitor).
z
Input the nominal value via the numeral keys (5 digits with decimal
maximum).
z
Press ↵ key
z
After the BAR at the bottom of LCD monitor is filled fully, the “Nominal
Value” input is done.
SPEED
SLOW
L 1.2345H
Q 0.6789
DISPLAY
VALUE
MODE
L/Q
TESTING
CIRCUIT
SERIES
F : 1.000 kHz R.H OFF
V : 1.000 V
C.V OFF
AUTO MANU INT.B OFF
MENU
CAP. R/L OFFSET
OFFSET
SET SORT
SORT
4-4-5. Selection of Measurement Speed
SET PARAMETER
SETTING
One of three measurement speeds SLOW, MEDIUM, or FAST could be
selected (Figure 4-7). The continuous mode speeds are about 1, 5, and 12
measurement per second respectively. The trade-off is accuracy vs. speed. LCR817/819 will take a more accurate measurement at a slower rate. The trade-off is
as follows
CALIBRATION
CALBRAT
SLOW speed : More than 1 measurement per second, at 0.05% accuracy
(or better)
MEDIUM speed : More than 3 measurements per second, at 0.1% accuracy
(or better)
Press menu
key
Press F2 key
to the sort
menu
EXIT
NOM.VAL
24.870
NOM.VAL = 77.000pF
OPTION 1
Press F1 key to
input the
nominal value
HANDLER
FAST speed : More than 7 measurements per second, at 0.24% accuracy
(or better)
* For the details of accuracy, please refer to the specifications.
* Regarding the models LCR-826/827/829, please refer to 4-5-10 Handler
Interface Timing.
16
EXIT
Figure 4-6. Steps of “Nominal Value” input.
17
LCR-800 Series User Manual
LCR-800 Series User Manual
Press F1 key to select the three different measurement
speed. (SLOW, MEDIUM, or FAST)
L 1.2345H
Q 0.6789
TESTING
F : 1.000 kHz R.H OFF
V : 1.000 V
C.V OFF
AUTO MANU INT.B OFF
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
Press the numerical 7 key
to select either internal or
external bias voltage
7
8
BIAS
ON/OFF
4
5
6
2
3
.
_
PPM
1
C.V
Figure 4-7. Selection of measurement speed
0
4-5. Measurement Conditions
Press the numerical 8 key to
turn either internal or external
bias voltage on or off.
R.H
S
T
A
R
T
9
FREQ
4-5-1. Bias Voltage
There are two available bias voltage modes: “Internal” and “External”.
Internal:
An internal DC 2 volts bias voltage will apply to the device under test.
External:
An external DC bias voltage between 0 and 30 volts can be applied to the device
L 1.2345H
Q 0.6789
under test. The external bias connection is located on the rear panel. The
maximum current is 200mA. The supply of bias voltage has to be floating, don’t
connect either side to ground. It’s better to wait approximately 1 second for
taking a reading after initiating a testing process, therefore, the device under test
will stabilize after bias voltage applied. The DC bias voltage should be applied
only to capacitors in general. If the DC bias voltage is applied to device of low
impedance, the unreliable testing results will occur. When the external bias is
applied, please join the function of C.V.ON (Constant voltage On) in use.
18
TESTING
F : 1.000 kHz R.H OFF
V : 1.000 V
C.V OFF
AUTO MANU INT.B OFF
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
Indication of internal or
external bias voltage
Figure 4-8. Selection of “BIAS” voltage
19
LCR-800 Series User Manual
LCR-800 Series User Manual
Steps of “BIAS” voltage selection (Figure 4-8):
z
Press compound key 7 to selects the “INTERNAL BIAS” or “EXTERNAL
BIAS” on the main menu. (If this key function is switched to “INTERNAL
BIAS”, the bottom of LCD monitor will display the “INT.B” message. If the
external DC bias is selected, the bottom of LCD monitor will display the
“EXT.B” message.
z
Press compound key 8 to turn either “INTERNAL BIAS” or “EXTERNAL
BIAS” mode “ON” or “OFF” on the main menu.
4-5-2. Test Frequency
The numerical input of test frequency accepts up to 5 digits with decimal. User
can input any desired frequencies, however, the actual frequency executed for the
LCR Meters is always the closest one of the 504 available frequencies. The 504
frequencies can be calculated by the following formulas:
3kHz/n,
60kHz/n,
200kHz/n,
where n range is from 13 to 250 (freq. 0.012 to 0.23077kHz)
where n range is from 4 to 256 (freq. 0.23438 to 15kHz)
where n range is from 1 to 13 (freq. 15.385 to 200kHz)
The “nominal value” of an available frequency can be calculated from the proper
one of the three formulas.
The range of test frequency between 12Hz and 10kHz is for LCR-817/827,
between 12Hz and 100kHz for LCR-819/829, between 100Hz and 2kHz for
LCR-816/826, and between 12Hz and 200kHz for LCR-821. To select the test
frequency, just key in the desired frequency via these numerical keys, the LCR
Meters will take the nearest available test frequency from the 504 available test
frequencies automatically.
Steps of “Test Frequency” selection (Figure 4-9):
Press compound key FREQ .
z
Input the desired frequency in kilohertz.
Press ↵ key.
Note: After test frequency has been changed, the zeroing of “Open/Short
circuit” must be done again for the best accuracy.
z
7
8
BIAS
ON/OFF
4
5
6
2
3
.
_
PPM
1
C.V
0
R.H
S
T
A
R
T
9
FREQ
Press - key to input test
frequency
Press enter key to confirm the
inputs.
L 1.2345H
Q 0.6789
TESTING
F : 1.000 kHz R.H OFF
C.V OFF
V : 1.000 V
AUTO MANU INT.B OFF
SPEED
SLOW
DISPLAY
VALUE
MODE
L/Q
CIRCUIT
SERIES
MENU
z
20
test frequency
Figure 4-9. Inputs of test frequency
21
LCR-800 Series User Manual
LCR-800 Series User Manual
4-5-3. D/Q in PPM (parts per million)(Invalid for LCR-816/826)
If the value of D or Q is less than 0.0100, user can select DQ in PPM to improve
the resolution by a factor of 100. The units of D and Q in PPM are dimensionless
and expressed as a decimal ratio with the multiplier of 1000000. User can just
SPEED
SLOW
L 1.2345H
Q 0.6789
DISPLAY
VALUE
MODE
L/Q
press compound key 4 to select the unit of D or Q in PPM. To disable the DQ in
TESTING
PPM feature, press the same key again.
4-5-4. Test Voltage
The range of test voltage is from 5mV to 1.275V in increments of 5mV. The
actual voltage through the DUT is never more than the source voltage. The DUT
voltage in general. The programming of test voltage is as follows (Figure 4-10):
MENU
CAP. R/L OFFSET
OFFSET
z
Press MENU key.
SET SORT
z
Press F3 key to select “SETTING” menu.
SET PARAMETER
z
Press F2 key to select “VOLT” menu.
SETTING
z
Input the desired value via the numeral keys.
CALIBRATION
CALBRAT
z
z
Press menu
key
Test Voltage
impedance and the source resistance of the LCR Meters decide the actual test
voltage. The smallest voltage through the DUT will be 20% smaller of the source
CIRCUIT
SERIES
F : 1.000 kHz R.H OFF
V : 1.000 V
C.V OFF
AUTO MANU INT.B OFF
SORT
Press ↵ key
After the BAR at the bottom of LCD monitor is filled fully, the “Test
Press F3 key
to the setting
menu
EXIT
Voltage” input is done.
Note: When the test frequency is at 200kHz, the test voltage must be more than
100mV.
MOMORY NO: 1
MOMORY
1
If the DUT have to be measured at a particular test voltage, the LCR Meters
VOLTAGE = 1.000 V
VOLT
1.000
provides the constant voltage feature. After “Constant Voltage” is selected, The
AVERAGE = 1
4-5-5. Constant Voltage Source
AVGE
1
LCR Meters will keep a source resistance of 25Ω. Therefore, the test voltage is
Press F2 key to
input the desired
test voltage
constant for any DUT impedance which large than 25Ω. If “Constant Voltage” is
selected, the measurement accuracy will cause a reduction by a factor of three.
EXIT
User can just press compound key 1 to select the feature of “Constant Voltage”.
To disable this feature, press the same key again.
22
Figure 4-10. Programming of test voltage.
23
LCR-800 Series User Manual
LCR-800 Series User Manual
4-5-6. Range Hold
If a DUT is removed from the test cable or fixture during the “Continuous” mode,
the feature of “Range Hold” can avoid range switching. Due to the test time can
be reduced, “Range Hold” is a useful utility for repetitive devices measurement.
L 1.2345H
Q 0.6789
SPEED
SLOW
DISPLAY
VALUE
User can just press compound key 0 to select the feature of “Range Hold”. To
TESTING
disable this feature, press the same key again.
4-5-7. Averaging
If this function is enable, the testing time will multiple by the number of tests
MODE
L/Q
F : 1.000 kHz R.H OFF
V : 1.000 V
C.V OFF
AUTO MANU INT.B OFF
CIRCUIT
SERIES
Press menu
key
MENU
(from 1 to 255). Hence, the accuracy can be enhanced. The measurement time
will be increased to the contrary. The programming of “Averaging” is as follows
(Figure 4-11):
CAP. R/L OFFSET
z
Press MENU key.
z
Press F3 key to select “SETTING” menu.
z
Press F3 key to select “AVGE” menu.
z
Input the desired value via the numeral keys.
z
z
Press ↵ key
After the BAR at the bottom of LCD monitor is filled to the full, the
OFFSET
SET SORT
SORT
SET PARAMETER
SETTING
CALIBRATION
CALBRAT
Press F3 key
to the setting
menu
EXIT
“Averaging” input is done.
MOMORY NO: 1
MOMORY
1
VOLTAGE = 1.000 V
VOLT
1.000
AVGE
1
AVERAGE = 1
Press F3 key to
input the desired
averaging amount
EXIT
Figure 4-11. Programming of averaging.
24
25
LCR-800 Series User Manual
LCR-800 Series User Manual
4-5-8. Memory
The LCR Meters have two memory functions: Recall and Store. The current
measurement conditions can be saved into the memory unit or recalled a
previously saved measurement conditions set from the LCR Meters memory.
L 1.2345H
Q 0.6789
SPEED
SLOW
DISPLAY
VALUE
There are 100 memory blocks totally. The programming of “Memory Store /
Recall” can be used as described below (Figure 4-12):
z
Press MENU key.
z
Press F3 key to select “SETTING” menu.
z
Press F1 key to select “MEMORY” menu.
z
Press compound key 1 to recall a previously saved memory block. or
z
Press compound key 2 to store the current measurement conditions into
z
z
z
F : 1.000 kHz R.H OFF
V : 1.000 V
C.V OFF
AUTO MANU INT.B OFF
memory.
CAP. R/L OFFSET
Input the number of desired memory block. (1~100)
SET SORT
Press ↵ key
After the BAR at the bottom of LCD monitor is filled to the full, the process
of “Memory Store/Recall” is done.
MODE
L/Q
TESTING
CIRCUIT
SERIES
Press menu
key
MENU
OFFSET
SORT
SET PARAMETER
SETTING
CALIBRATION
CALBRAT
Remark: For LCR-827/829, the function of Recall and Store is not only used in
Press F3 key
to the setting
menu
EXIT
the normal memory status, also it can be extended to sorting and BINSUM setting.
MOMORY NO: 1
MOMORY
1
VOLTAGE = 1.000 V
VOLT
1.000
Press F1 key to
select the memory
store/recall function
AVGE
1
AVERAGE = 1
EXIT
Figure 4-12. Programming of memory store/recall.
26
27
LCR-800 Series User Manual
LCR-800 Series User Manual
Press compound key 1 to
select the memory recall
function
Press compound key 2 to
select the memory store
function
(1)RECALL:(2)STORE
MOMORY
1
VOLT
1.000
VOLTAGE = 1.000 V
AVGE
1
AVERAGE = 1
EXIT
Input number of desired memory block for memory recall function
MOMORY
1
RECALL NO:
VOLTAGE = 1.000 V
VOLT
1.000
AVERAGE = 1
AVGE
1
EXIT
Input number of desired memory block for memory store function
MOMORY
1
STORE NO:
4-5-9. Handler Interface(Only for LCR-826/827/829)
FUNCTION
Start Measurement(I)
End of test(O)
Data acquisition over , DUT removal OK.(O)
RLC FAIL(O)
No-Go , D or Q Fail
Go , BIN 1 Judgement
Go , BIN 2 Judgement
Go , BIN 3 Judgement
Go , BIN 4 Judgement
Go , BIN 5 Judgement
Go , BIN 6 Judgement
Go , BIN 7 Judgement
Go , BIN 8 Judgement
Go , BIN 9 Judgement
Go , BIN 10 Judgement
Go , BIN 11 Judgement
Go , BIN 12 Judgement
Go , BIN 13 Judgement
Panel Lock
GND
VCC
VOLT
1.000
VOLTAGE = 1.000 V
AVGE
1
AVERAGE = 1
EXIT
Figure 4-12. Programming of memory store/recall. (Cont.)
28
29
HANDLER
INTERFACE
Signal Name
Pin NO.
/I_E_TRIG
24
/O_EOM
23
/O_INDEX
22
/O_P_HI
19
/O_P_LO
20
/O_P_OVER
15
/O_S_REJ
21
/O_S_OVER
14
/O_BIN_ 1
1
/O_BIN _2
2
/O_BIN_ 3
3
/O_BIN_ 4
4
/O_BIN_ 5
5
/O_BIN_ 6
6
/O_BIN_ 7
7
/O_BIN_ 8
8
/O_BIN_ 9
9
/O_BIN_ 10
10
/O_BIN_ 11
11
/O_BIN_ 12
12
/O_BIN_ 13
13
/I_K_LOCK
25
GND
16,18
VCC
17
LCR-800 Series User Manual
LCR-800 Series User Manual
z
Handler Interface Typical Electrical Characteristics
Voltage Output Rating
Comparison Signals
Maximum Current
Low
High
/BIN1-/BIN9
/AUX_BIN
*
*
/OUT_OF_BIN
/PHI
5mA
≦0.5V +5V~+24V
/PLO
Control Signals
/INDEX
/EOM
* R408~R427 Pull-high Resistor need to be changed.
Handler Interface Typical Electrical Characteristics
Input Current (Low)
Input
Pull-high Voltage
Comparison Signals
Low
High
+5V
+12V
/EXT_TRIG
5mA
12mA
≦1V 5V~15V
/KEY_LOCK
5mA
12mA
≦1V 5V~15V
Measurement Complete signal:
/O_INDEX: This signal will go the low level when analog measurement is
completed and will go the high level when next trigger is
Note
active. The handler can then remove the DUT from the fixture
and replace it with another DUT during the time when
/O_INDEX is at low level.
z
Compare Output Signals:
/O_BIN_1- /O_BIN_13 : The GO judgment is active according to the
measured value judgment results (low level),
for example, if the comparator assigns a
component to BIN1, the /O_BIN_1 line will go
the low level till T4 (see Handler Interface
Note
Timing). /O_BIN_2 - /O_BIN_13 are all
inactive (high level).
/O_P_HI: Main parameter failure: When the main parameter is greater than
Max, it will go the low level till T4 (see Handler Interface
z
Timing).
Input Signals
/I_E_TRIG: This signal is a measurement start signal. To start
/O_P_LO: Main parameter failure: When the main parameter is lower than
measurement by setting low pulse to minimum 5us for
Min, it will go the low level till T4(see Handler Interface
/I_E_TRIG signal. This is a low edge active.
Timing).
/I_K_LOCK: This signal is to lock all front panel controls. The front panel
is disable when /I_K_LOCK is at low level, and enable
/O_P_OVER: It will go the low level when main parameter is greater than
Max or is lower than Min.
when /I_K_LOCK is at high level.
/O_S_REJ,/O_S_OVER: It will go the low level when secondary is greater
than D_Max (if MODE=C/D) or is lower than
D_Min.
30
31
LCR-800 Series User Manual
LCR-800 Series User Manual
4-5-10. Handler Interface Timing (only for LCR-826/827/829)
z
Judgment Complete Signal
/O_EOM: This signal becomes active (low level) when the measurement
and comparator judgment are completed, and all handler
interface output are stable. It will go the high level when next
trigger Signal (/I_E_TRIG) becomes active (low edge).
* DATA: /O_BIN_1-_BIN_13,/O_P_OVER,/O_S_OVER,/O_S_REJ,/O_P_HI,/O_P_LO
32
33
LCR-800 Series User Manual
LCR-800 Series User Manual
Time
Minimum Value Maximum Value
T1
Trigger Pulse Width
T2
Measurement Start Delay Time
T3
T4
5 us
—————
140 us
—————
/EOM Delay Time After Data Output
5 us
—————
Calculation and binning Time
6 ms
—————
Trigger Wait Time
After /EOM
Output (T6)
Analog Measurement Time (T5)
SPEED
0.012kHz 0.1kHz 0.12kHz 1kHz 10kHz 100kHz OFF BIN VALUE
SLOW
817ms
MEDIUM 817ms
FAST
817ms
901ms 901ms 903ms 873ms 873ms 2ms 4ms 16ms
125ms 105ms 59ms 53ms
53ms 2ms 4ms 16ms
125ms 103ms 27ms 17ms
17ms 2ms 4ms 16ms
ACCURACY
FAST
MODEL
LCR-827
LCR-829
LCR-826
MEDIUM
SLOW
5. SPECIFICATIONS
Measurement Parameters:
Inductance (Ls/Lp)*, Capacitance (Cs/Cp), Resistance (Rs/Rp), Dissipation (D),
Quality Factors (Q), Equivalent Series Resistance (ESR) and Equivalent Parallel
Resistance (EPR), Impedance (|Z|), Phase angle of Impedance [degree] (θ).
Measurement Models:
Six kinds of measurement modes can be selected. Two measurement parameters
measured and displayed simultaneously.
R/Q, C/D, C/R, L/Q, Z/θ, L/R.
Display Ranges:
Primary Display
Inductance (L)
Capacitance (C)
Resistance (R)
Absolute of Impedance (Z)
Secondary Display:
Dissipation factor (D)+
Quality factor (Q)**
: 0.00001mH
: 0.00001pF
: 0.00001Ω
: 0.00001Ω
~
~
~
~
99999H
99999μF
99999kΩ
99999kΩ
: 0.0001
: 0.0001
~ 9999
~ 9999
Phase angle of Impedance (degree) : -180.00∘ ~ 180.00∘
0.5%
0.2%
0.1%
Equivalent Series Resistance (ESR)+
Equivalent Parallel Resistance (EPR)+
Dissipation factor (D)+ in ppm
Quality factor (Q)** in ppm
DELTA %
: 0.0001Ω
: 0.0001Ω
: 1 ppm
: 1 ppm
: 0.00001%
~
~
~
~
~
9999 kΩ
9999 kΩ
9999 ppm
9999 ppm
99999%
If any of these quantities is negative, the “-“ negative indicator is displayed.
*s=series, p=parallel, ESR=Rs ** with L or R + with C
Note: Only LCR-821 has Z/θ and L/R measurement parameter.
34
35
LCR-800 Series User Manual
LCR-800 Series User Manual
Accuracy:
R, L, C, Z : 0.05%(Basic)+
D, Q
:.0005(Basic)+
θ
: 0.03∘(Basic)+
*LCR-816/826/827/829 is one fold less accuracy than LCR-817/819.
(LCR-816/826: 0.1%, LCR-827/829: 0.1%, LCR-817/819/821: 0.05%)
+: Please refer to page 39~43 for details.
Equivalent Circuit:
The L, C, or R equivalent SERIES or PARALLEL circuit can be selected by
keypad.
* When select Z/θ mode, the PARALLEL circuit is invalid. Please refer to 4-42. Series & Parallel Equivalent Circuit from page 12~13 for details.
Test Frequency:
There are 504 test frequencies between 12Hz and 200kHz that can be selected
by using keypad for LCR-821. The frequency range of LCR-819/829 covers
from 12Hz to 100kHz (total test frequencies: 503 points), LCR-817/827 covers
from 12Hz to 10kHz (total test frequencies: 489 points), LCR-816/826 covers
from 100Hz to 2kHz (total test frequencies: 16 points).
Measurement Displays:
The measured results can be shown on the LCD monitor in three ways:
: The measured quantities of R/Q, C/D, C/R, L/Q, Z/θ, L/R.
*The resolution of primary display (L, C, R or Z) is to be five
digits.
*The resolution of secondary display (D, Q, R with C, or R with
L) is to be four digits.
*The resolution of secondary display (θ) is to be2 digits after
decimal dot.
2. DELTA % : The DELTA% shows the percent deviation of the measured L,
C, R or Z value from a saved NOMINAL VALUE. The sign of
deviation is indicated.
3. DELTA : The LCR difference is similar to the DELTA% except that the
deviation is shown in suitable units (ohms, henries, etc.)
1. VALUE
Measurement Speed (LCR-816/817/819/821):
Measurement Modes:
Two modes are available: AUTO and MANUAL.
“AUTO” mode is measuring continuously, updating the display after each
measurement.
“MANUAL” mode is activated by the START keypad, the measured result is
holding on the LCD monitor until next measurement started.
Average:
The AVERAGE of any number of measurements from 1 to 255 can be made as
desired in either of the two measurement modes.
Test Voltage:
The test voltage range for LCR-817/819/827/829/821 is from 5mV to 1.275V,
each step is 5mV, for LCR-816/826 is from 0.1V to 1.275V, each step is 5mV.
Note: When the test frequency is at 200kHz, the test voltage must be greater
than 100mV.
Memory:
100 memory blocks totally.
DC Bias:
A 2V internal bias can apply to capacitors during measurement.
Up to 30VDC external bias can apply to capacitors during measurement via the
two terminals (located on the rear panel). The applied current should not over
200mA.
* The external bias specified on the rear panel is at 30VDC, actually, it can be
tolerated to 35VDC.
SLOW : 896ms
MEDI : 286ms
FAST : 135ms
*For LCR-826/827/829, please refer to 4-5-10. Handler Interface Timing.
36
37
LCR-800 Series User Manual
LCR-800 Series User Manual
The Error value of LCR-817/819/821 (Double the error value for LCRLCD monitor:
240×128 dot matrix C.C.F.L. back light LCD with adjustable contrast.
827/829/816/826)
Battery:
DC source for system memory and calibration data backup, a replaceable 3V
lithium battery (BR-2/3A type), which probably needs to be replaced after using
for 3 years.
Every time when the battery is replaced, the LCR Meters should be recalibrated.
C: 2 counts±0.03%+0.02%[(1+Ka)# or (X/Ymax )# or (Ymin/X)# ] (1+│D│)(1+Kb+Kc)
Operation Environment:
Indoor use,
Altitude up to 2000M
Installation Category II
Pollution Degree
2
Operating temperature: 10℃~50℃, < 85% relative humidity
Storage temperature:-20℃~60℃
z
R: 2 counts±0.03%+0.02%[(1+Ka)# or (X/Ymax )# or (Ymin/X)# ] (1+│Q│)(1+Kb+Kc)
L: 2 counts±0.03%+0.02%[(1+Ka)#or(X/Ymax)#or(Ymin/X)# ] (1+ 1 /│Q│)(1+Kb+Kc)
|Z|: Ze = Treat R, L or C as the object under test conforming above formula:
example:
When the object under test is C, select:
Ze = 2 counts±0.03%+0.02%[(1+Ka)#or(X/Ymax)#or(Ymin/X)# ] (1+│D│)(1+Kb+Kc)
When the object under test is R, select:
Ze = 2 counts±0.03%+0.02%[(1+Ka)#or(X/Ymax)#or(Ymin/X)# ] (1+│Q│)(1+Kb+Kc)
When the object under test is L , select:
Ze=2 counts±0.03%+0.02%[(1+Ka)#or(X/Ymax)#or(Ymin/X)#](1+ 1 /│Q│)(1+Kb+Kc)
z
AC Power Source:
AC 100~240V, 50Hz~60Hz
The formula for primary readout accuracy of C, R, L, and Z.
The formulas of secondary readout accuracy for D, Q, and θ.
Error
Power Consumption:
45 Watts maximum
D with C 2count±0.0003 + 0.0002[(1+Ka)# or (X/Ymax) # or (Ymin/X) #] (1+│D│+D²)(1+Kb+Kc)
Fuse Replacement:
Slow-Blow, 3A, 250V
Q with L 2count±0.0003 + 0.0002[(1+Ka) # or (X/Ymax) # or (Ymin/X) #] (1+│Q│+Q²)(1+Kb+Kc)
Q with R 2count±0.0003 + 0.0002[(1+Ka) # or (X/Ymax) # or (Ymin/X) #] (1+│Q│+Q²)(1+Kb+Kc)
Dimensions:
330mm (W) × 149mm (H) × 437mm (D)
Weight:
5.5 kg
θ with Z θe=(180/π) x (Ze/100)
# 1. if X≧Ymax, please select (X/Ymax)
2. if X≦Ymin, please select (Ymin/X)
3. if Ymin<X<Ymax, please select (1+Ka)
4. Ze is impedance error
5. θe is θ error
38
39
LCR-800 Series User Manual
LCR-800 Series User Manual
z
The formulas for secondary readout accuracy of R with C.
Error
D≧1
2count + 0.02%[(1+Ka)* or (Rx/Rmax)* or (Rmin/Rx)*] (1+ 1/│D│)(1+Kb+Kc)+0.03%
D≦1
2count + 0.02%[(1+Ka)** or (Cx/Cmax)** or (Cmin/Cx)**] (1+ 1/│D│)(1+Kb+Kc)+0.03%
z
The formulas for secondary readout accuracy of R with L.
Error
Q≦1
2count + 0.02%[(1+Ka)* or (Rx/Rmax)* or (Rmin/Rx)*] (1+ | Q│)(1+Kb+Kc)+0.03%
Q≧1
2count + 0.02%[(1+Ka)** or (Lx/Lmax)** or (Lmin/Lx)**] (1+| Q│)(1+Kb+Kc)+0.03%
*: 1. If Rx≧Rmax, please select (Rx/Rmax)
2. if Rx≦Rmin, please select (Rmin/Rx)
3. if Rmin<Rx<Rmax, please select (1+Ka)
**: 1. If Cx≧Cmax, please select (Cx/Cmax)
2. if Cx≦Cmin, please select (Cmin/Cx)
3. if Cmin<Cx<Cmax, please select (1+Ka)
Rx and Cx are value of the component being tested.
Rmax, Rmin, Cmax and Cmin are ranges constant (refer to Table B).
Table A: (for range 1,2,3) –Kc
Voltage
0.03≦V<0.1 0.1≦V<0.25 0.25≦V<1
Frequency
35
12
9
0.012≦F<0.03
30
8
5
0.030≦F<0.1
25
6
3
0.1≦F<0.25
20
5
2
0.25≦F<1
1
14
4
1
15
5
2
1<F≦3
15
6
3
3<F≦6
15
8
5
6<F≦10
20
10
6
10<F≦20
30
22
18
20<F≦50
50
40
35
50<F≦100
200
Nonuse
80
50
F: test frequency in kHz
Where
Ka: Constant Voltage factor
Constant Voltage On , Ka = 2
Constant Voltage Off, Ka = 0
Kb: Test Speed factor
Speed = SLOW
, Kb = 0
Speed = MEDIUM , Kb = 3
Speed = FAST
,Kb = 10
Kc: Frequency & RMS Voltage factor (refer to Table A)
X: X is value of the component being tested.
Y: Y is range constant (refer to Table B)
40
41
1≦V≦1.265
7
3
2
1
0
1
2
3
5
15
30
45
LCR-800 Series User Manual
LCR-800 Series User Manual
Example of C-D Accuracy Calculation
Table A: (for range 4)-Kc
Measurement Conditions
Voltage
Frequency
0.012≦F<0.03
0.030≦F<0.1
0.1≦F<0.25
0.25≦F<1
1
1<F≦3
3<F≦6
6<F≦10
10<F≦20
20<F≦50
50<F≦200
0.03≦V<0.1 0.1≦V<0.25
70
50
35
25
15
17
25
60
Not specified
0.25≦V<1
1≦V≦1.265
Frequency: 1 kHz
10
6
4
2
1
3
10
20
65
7
3
2
1
0
2
6
15
50
Test Voltage: 1V
20
13
9
6
4
6
15
30
100
This range is not used above 20kHz
Test Speed=SLOW
Constant Voltage=OFF
R.H=OFF
C measure: 1 μF
D measure: 0.0009
Then:
C Error = 0.03% + [0.02% x (1+ Ka) x (1+0.0009) x (1+Kb+Kc)]
=0.03%+[0.02% x (1+0) x (1.0009) x (1+0+0)]
F: test frequency in kHz
=0.05%±2 counts
Table B-1: Range Hold
D Error = 0.0003+[0.0002 x (1+Ka) x (1+0.0009+0.0009²) x (1+Kb+Kc)]
Component
Inductor
Capacitor
Resistor/Impedance
=0.0003+[0.0002 x (1+Ka) x (1+0.0009+(0.0009²) x (1+Kb+Kc)]
Max
Min
Max
Min
Max
Min
Range1
16mH /f
1mH /f
25uF /f
1.6uF /f
100Ω
6.25Ω
Range2
256mH /f
16mH /f
1600nF /f
100nF /f
1.6kΩ
0.1kΩ
Z Error = C Error = 0.05%
Range3
4100mH /f 256mH /f
100nF /f
6.4nF /f
25.6kΩ
1.6kΩ
θ Error =(180/π) x (Ze/100) =57.3∘x (0.05/100)=0.03∘
6400pF /f
400pF /f
410kΩ
25.6kΩ
Range
Range4*
65H/f
4.1H /f
=0.0003 + [0.0002 x (1+0) x (1.0009) x (1+0+0)]
=0.0005±2 counts
f = test frequency in kHz
*: This range is not used above 20kHz
Table B-2: Auto Range
Component
Range
Auto range
Inductor
Capacitor
Resistor/Impedance
Max
Min
Max
Min
Max
Min
65H /f**
1mH /f
25uF /f
400pF /f **
410kΩ**
6.25Ω**
**: Above 20kHz, Cmin = 6.4 nF/f ,and Lmax = 4100mH/f
f = test frequency in kHz.
42
43
LCR-800 Series User Manual
LCR-800 Series User Manual
6. MESSAGE CODE
This section describes the message code for the LCR Meters.
OVER-01
If the impedance of “Device-under-test” is small than the existing
measurement range of the LCR Meters, the “OVER-01” message will be
displayed on the LCD monitor.
To clear the LCR Meters, use soft cloth dampened in a solution of mild detergent
Do not use chemicals containing benzine, benzene, xylene, acetone, toluene, or
Inductance: XL=2πfL
similar solvents.
where f=test frequency in Hz
If the inductor of “Device-under-test” is very large at a very high “test
Do not use abrasive cleaners on any portion of this equipment.
frequency”, the “Resonance effect” will occur and the impedance will
7-2. Battery Replacement
decrease. Hence, the measured value is useless. Meanwhile, an “OVER-01”
A replaceable 3V lithium battery (BR-2/3A type) supplies the backup power of
message will be displayed on the LCD monitor.
non-volatile memory for the LCR Meters. This battery probably needs to be
replaced after using for 3 years.
Solution:
Turn on the “Constant Voltage” mode (please refer to 4-5-5. Constant
Voltage Source, page 22, for details).
2.
LCR Meters before attempting to clean the instrument.
into the cabinet and cause damage.
Capacitance: XC=1/2πfC
1.
Remove the AC input power (disconnect and remove the power cord) from the
and water. Do not spray cleaner directly onto the instrument, since it may leak
Calculation formula:
2.
This section includes the basic maintenance information for the LCR Meters.
7-1. Cleaning
Cause:
1.
7. MAINTENANCE
Select the less measurement range. Please refer to table B-1: Range Hold,
page 42, in order to reach the exact measurement range. Turn on the “Range
Hold” mode after the measurement range fixed.
CAUTION: Danger of explosion if battery is incorrectly replaced.
Replace only with the same or equivalent type recommended by
the manufacturer. Discard used batteries according to the
manufacturer’s instructions.
Note: Both two solutions will reduce the accuracy of the LCR Meters.
44
45
LCR-800 Series User Manual
LCR-800 Series User Manual
7-3. Recall Calibration
When the measured value is inaccurate, you can use the Recall Calibration
CAP. C/D OFFSET
function to rework by proceeding the steps as follows:
SET SORT
SORT
z
Press MENU key. (Figure 7-1)
SET PARAMETER
z
Press F3 key to select “SET PARAMETER” function.(Figure 7-2)
CALIBRATION
z
Press F4 key to select “RECALL CALIBRATION” function.(Figure 7-3)
z
Press compound Key 1 to select the “Yes” to start recall calibration.
SETTING
CALBRAT
Press F3 key
to the setting
menu
EXIT
(Figure 7-4)
z
OFFSET
Figure 7-2
Press compound Key 2 to select the “No” to stop recall calibration.
(Figure 7-4)
※ If the function keys are not active, do not power off and wait another 2
minutes, then try again.
SPEED
SLOW
C 13.450uF
D .1652
MEMORY
0
77.000
MEMMORY NO: 0
VOLTAGE=
1.000
AVERAGE=
1
V
NOM.VAL
1
RECALL
RECALL CALIBRATION
DISPLAY
VALUE
EXIT
MODE
C/D
TESTING
CIRCUIT
SERIES
F :1.0000 kHz R.H OFF
V :1.000 V
C.V OFF
AUTO MANU INT.B OFF
Figure 7-3
Press menu
key
MEMORY
0
77.000
MEMMORY NO: 0
MENU
Figure 7-1
VOLT
1.000
VOLTAGE=
1.000
AVERAGE=
1
VOLT
1.000
V
NOM.VAL
1
RECALL
RECALL CALIBRATION
Are you sure?(1)Yes
(2)No
EXIT
Figure 7-4
46
47
Press F4 key to
input the recall
function
LCR-800 Series User Manual
LCR-800 Series User Manual
8. OPTION 1(BIN Functions, only for LCR-826/827/829)
8-1. BIN FUNTIONS for Components Sorting:
C 13.450uF
D .1652
SPEED
SLOW
DISPLAY
VALUE
Proceed BIN FUNTIONS steps(Figure 8-1):
z
Press MENU key.
z
Press F2 key to select “SORT” function.
z
Press F2 key to select “OPTION1 ” function.
MODE
C/D
TESTING
F :1.0000 kHz R.H OFF
V :1.000 V
C.V OFF
AUTO
MANU INT.B OFF
CAP. C/D OFFSET
CIRCUIT
SERIES
Press menu
key
MENU
OFFSET
SET SORT
SORT
SET PARAMETER
SETTING
CALIBRATION
CALBRAT
Press F2 key
to the sort
menu
EXIT
NOM.VAL = 77.000pF
OPTION1
NOM.VAL
77.000
HANDLER
Press F2 key to
input the
option1
function
EXIT
Figure 8-1.: OPTION 1(BIN FUNTIONS)setting
48
49
LCR-800 Series User Manual
LCR-800 Series User Manual
8-2. BIN setting conditions:
SPEED setting:
z
Press F1/F2 key to select the white inverted characters “SPEED”.
z
Press F3 key to select FAST, SLOW or MEDI shown as Figure 8-4 below.
For further details, please refer to 4-4-5.
MODE setting:
z
Press F1/F2 key to select the white inverted characters “MODE”.
z
Press F3 key to select C/D or C/R or L/Q or R/Q shown as Figure 8-2 below.
For further detailed description, please refer to 4-4-1.
SET BIN
SET BIN
BIN SUM
Press F1/F2
key to the
mode item
MODE :C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C
D
SET
BIN
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
Press F3 key
to the
C/R.L/Q.R/Q
CIRCUIT setting:
z
Press F1/F2 key to select the white inverted characters “CIRCUIT”.
z
Press F3 key to select SER or PAR shown as Figure 8-3 below. For further
details, please refer to 4-4-2
Press F1/F2 key
to the
CIRCUIT
MODE:C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
SET
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
EXIT
D
SET
BIN
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
EXIT
SET BIN
Press F3 key
to the SER.
PAR
BIN SUM
50
Press F1/F2
key to the
DISPLAY
MODE:C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY :BIN
C
D
SET
BIN
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
EXIT
Figure 8-5
Figure 8-3
Press F3 key
to the
FAST/SLOW/MEDI
DISPLAY setting:
z
Press F1/F2 key to select the white inverted characters “DISPLAY”.
z
Press F3 key to select BIN, VAL or OFF shown as Figure 8-5 (When the
BIN is selected, BIN1~BIN13 represents PASS, while PHI, PLO, and SREJ
represents FAIL result. The VAL is to display test value while the OFF
represents as non-display.)
※The display speed for above three selections is different from one another.
D
BIN
Press F1/ F2
key to the
SPEED
Figure 8-4
Figure 8-2
C
C
EXIT
SET BIN BIN SUM
BIN SUM
MODE :C/D CIRCUIT:SER.
SPEED :SLOW DISPLAY:BIN
51
Press F3 key
to the
VAL.OFF
LCR-800 Series User Manual
LCR-800 Series User Manual
F setting (FREQUENCY):
INT.B setting:
z
Press F1/F2 key to select the white inverted characters “F”.
z
Press F1/F2 key to select the white inverted characters “INT.B”.
z
Key-in testing frequency through panel keyboard.
z
Press F3 or F4 key to select INT or ENT shown as Figure 8-8 below. For
z
Press ↵ key shown as Figure 8-6 below. For further details, please refer to
further details, please refer to 4-5-1.
4-5-2.
SET BIN
SET BIN
BIN SUM
Press F1/F2
key to the F
MODE:C/D
CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C
D
CLR
BIN
F:1.0000kHz MANU
V:1.000 V INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
BIN SUM
Press F1/F2 key
to the
INT.B
MODE:C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C
Press F3 key
to delete the
values input
in error
D
INT
BIN
Press F3/ F4
key to select
INT or .EXT
F:1.0000kHz MANU
V:1.000 V INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
EXIT
EXT
EXIT
Figure 8-8
Figure 8-6
INT.B ON OFF setting:
V setting (VOLTAGE):
z
Press F1/F2 key to select the white inverted characters “V”.
z
Press F1/F2 key to select the white inverted characters “ON/OFF” of INT.B.
z
Key-in testing voltage through panel keyboard.
z
Press F3 or F4 key to select ON or OFF shown as Figure 8-9 below.
z
Press ↵ key shown as Figure 8-7 below. For further details, please refer to
4-5-4.
SET BIN
BIN SUM
Press F1/F2 key
MODE:C/D
CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C
SET BIN
BIN SUM
Press F1/F2
key to the V
MODE:C/D
CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C
D
CLR
BIN
F:1.0000kHz MANU
INT.B OFF
V:1.000 V
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
Press F3 key
to delete the
values input
in error
ON
BIN
Press F3/ F4
key to select
ON or OFF
F:1.0000kHz MANU
V:1.000 V
INT.B: OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
OFF
EXIT
Figure 8-9
EXIT
Figure 8-7
52
to the OFF
D
53
LCR-800 Series User Manual
LCR-800 Series User Manual
RANGE setting:
DELAY setting:
z
Press F1/F2 key to select the white inverted characters “RANGE”.
z
Press F1/F2 key to select the white inverted characters “DELAY”.
z
PressF3 or F4 key to select UP or DOWN value within 1~4 range shown as
z
Key-in Delay value(0~99999ms) through panel keyboard.
Figure 8-10 below.
z
Press ↵ key. Please refer to Figure 8-12 below.
SET BIN
BIN SUM
C
SET BIN
Press F1/F2 key
to the RANGE
MODE:C/D
CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
D
C
UP
BIN
BIN SUM
Press F1/F2 key
to the DELAY
MODE:C/D CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
Press F3/F4
key to select
D
CLR
BIN
UP or DOWN
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
DOWN
EXIT
AVG setting:
C.V setting:
z
Press F1/F2 key to select the white inverted characters “C.V”.
z
Press F3 or F4 key to select ON or OFF shown as Figure 8-11 below. For
further details, please refer to 4-5-5.
z
Press F1/F2 key to select the white inverted characters “AVG”.
z
Key-in average value through panel keyboard.
z
Press ↵ key shown as Figure 8-13 below. For further details, please refer to
4-5-7.
SET BIN
BIN SUM
Press F1/F2 key
to the C.V
MODE:C/D
CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C
EXIT
Figure 8-12
Figure 8-10
SET BIN
BIN SUM
Press F1/F2 key
to the AVG
MODE:C/D
CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
C
D
D
ON
BIN
Press F3/F4
key to select
ON or OFF
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
C.V: OFF
RANGE:1
DELAY:00000mS AVG : 1
OFF
EXIT
CLR
BIN
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG: 1
EXIT
Figure 8-13
Figure 8-11
54
Press F3 key
to delete the
values input
in error
8-3. BIN Range Setting: Component Sorting Range
55
Press F3 key
to delete the
values input
in error
LCR-800 Series User Manual
LCR-800 Series User Manual
The step for BIN setting: (Please refer to Figure 8-14)
z
When the SET BIN inverts to white character, press F3 key.
z
Press F3 key to select C or D sorting (The selected is a primary parameter,
SET BIN
C
the other one is an assistant parameter.)
z
BIN clear setting by pressing F1 key.
z
Press F1 key to maintain setting, Press F2 key to delete the setting.
BIN SUM
MODE:C/D
CIRCUIT:SER.
SPEED:SLOW DISPLAY:BIN
D
SET
BIN
F:1.0000kHz MANU
V:1.000 V
INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
EXIT
BIN1
SORT BY C TOt_Bin:13
C_Nom.Val: 15.000pF
BIN
CLR
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
SORT
D
NEXT
BIN
D_Max:1.000
D_Min:.0010
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
NO
YES
SORT
D
NEXT
BIN
EXIT
Figure 8-14
56
Press F1 key
for clear
setting
Press F3 key
to sort C or D
EXIT
BIN1
SORT BY C TOt_Bin:13
C_Nom.Val: 15.000pF
D_Max:1.000
D_Min:.0010
Press F3 key
to the bin
setting
57
Press F1 key
for non-clear
Press F2 key
for clear
setting
LCR-800 Series User Manual
LCR-800 Series User Manual
Tot_BIN setting:
Press F1/F2 key to select the white inverted characters “Tot_Bin”.
z
Key-in needed BIN value through panel keyboard.
z
Press ↵ key. Please refer to Figure 8-15 below.
Note: Tot_Bin setting: It can select ascending from 1 to 13 BIN the most.
For example: If BIN1 sorting range is at ±5%, then BIN2 sorting range must
be larger than ±5%, the rest can be done accordingly.
Max setting:
z
Press
key
to F1/F2
the
Tot Bin
BIN 1
SORT BY C Tot Bin:13
C_Nom.Val: 15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
CLR
NEXT
BIN
D_Max:1.000
D_Min:.0010
PressF3
can deletekey
the
values input
in error
C_Nom.Val setting:
z
Press F1/F2 key to select the white inverted characters “C_Nom.Val”.
z
Key-in Nominal Value through panel keyboard.
z
Press ↵ key. Please refer to Figure 8-16 below.
CLR
NEXT
BIN
EXIT
58
Key-in SORT Max value through panel keyboard.
z
Press ↵ key. Please refer to Figure 8-17 below.
※After the value is input, it will proceed conversion of +% automatically.
Press F1/F2
key to the
Max
BIN 1
SORT BY C TOt_Bin:13
C_Nom.Val:15.000pF
CLR
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
NEXT
BIN
Press F3 key
can delete
the values
input in error
EXIT
“+” setting:
z
Press F1/F2 key to select the white inverted characters “+”.
z
Key-in the largest percentage of SORT Nominal Value through the panel
keyboard.
z
Press ↵ key. Please refer to Figure 8-18 below.
※After the value is input, it will proceed conversion of Max automatically.
Press F1/F2
key to the
C Nom.Val
BIN 1
SORT BY C TOt_Bin:13
C_Nom.Val: 15.000pF
Figure 8-16
z
Figure 8-17
Figure 8-15
D_Max:1.000
D_Min:.0010
Press F1/F2 key to select the white inverted characters “Max”.
D_Max:1.0000
D_Min:0.00010
EXIT
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
z
Press F3 key
can delete the
values input
in error
Press F1/F2
key to the
+
BIN1
SORT BY CTOt_Bin:13
C_Nom.Val: 15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max:1.0000
D_Min:0.00010
CLR
Press F3 key
to delete the
values input
NEXT
BIN
in error
EXIT
Figure 8-18
59
LCR-800 Series User Manual
LCR-800 Series User Manual
Min setting:
Press F1/F2 key to select the white inverted characters “Min”.
z
Key-in SORT Min value through panel keyboard.
z
Press ↵ key. Please refer to Figure 8-19 below.
※After the value is input, it will proceed conversion of -% automatically.
D_Max setting:
z
D_Max:1.0000
D_Min:0.00010
Press F1/F2 key to select the white inverted characters “D_Max”.
z
Set the largest D value of secondary parameter. Please refer to Figure 8-21
below.
CLR
PressF3 key
to delete the
values input
NEXT
BIN
in error
SORT BY C TOt_Bin:13
C_Nom.Val: 15.000pF
CLR
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
NEXT
BIN
D_Max:1.0000
D_Min :0.00010
EXIT
Figure 8-19
z
Press F1/F2 key to select the white inverted characters “-”.
z
Key-in the smallest minus percentage of SORT Nominal Value through the
panel keyboard.
Press ↵ key. Please refer to Figure 8-20 below.
EXIT
D_Min setting:
z
Press F1/F2 key to select the white inverted characters “D_Min”.
z
Set the smallest D value of secondary parameter. Please refer to Figure 8-22.
※After the value is input, it will proceed conversion of Min automatically.
Press F1/F2
key to the
-
BIN1
SORT BY CTOt_Bin:13
C_Nom.Val: 15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max:1.0000
D_Min:0.00010
CLR
NEXT
BIN
EXIT
Figure 8-20
60
Press F3 key
to delete the
values input
in error
Figure 8-21
“-” setting:
z
Press F1/F2
key to the
D_Max
BIN 1
PressF1/F2
key to the Min
BIN 1
SORT BY C TOt_Bin:13
C_Nom.Val:15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
z
Press F3 key
to delete the
values input
in error
Press F1/F2
key to the
D_Min
BIN 1
SORT BY C TOt_Bin:13
C_Nom.Val: 15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max :1.0000
D_Min:0.00010
CLR
NEXT
BIN
EXIT
Figure 8-22
61
Press F3 key
to delete the
values input
in error
LCR-800 Series User Manual
LCR-800 Series User Manual
The other BIN setting:
z
When the BIN 1 setting is finished, press F4 key to continue the BIN 2
setting until all the Tot BIN setting is completed (BIN1~BIN13 the most),
SET BIN
PressF1.F2
key to the
BIN SUM
BIN SUM
MODE:C/D
CIRCUIT:SER
SPEED:SLOW DISPLAY:BIN
C
please refer to Figure 8-23.
D
SET
BIN
F :1.0000kHz MANU
V :1.000 V INT.B OFF
RANGE:1
C.V : OFF
DELAY:00000mS AVG : 1
BIN
CLR
BIN 1
SORT BY C TOt_Bin:13
C_Nom.Val: 15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
SORT
D
NEXT
BIN
D_Max:1.0000
D_Min:0.00010
EXIT
Press F3 key
to the test
result
EXIT
Figure 8-24
Press F4 key
to the next
BIN
Figure 8-23
SORT BY C
NOM_VAL:15.000pF
D_MIN: .0010 D_MAX: 1.000
TEST RESULT
FAIL_ITEM
TOTAL
PHI ( C >MAX )
PLO ( C <MIN )
SREJ( D NG )
2
TOTAL
2
NEXT
PAGE
Press F4 to
CLR
CLR.
PASS_ITEM
BIN 1 - BIN 13
EXIT
TEST RESULT Display Summing-up:
z
Figure 8-25
Press F1/F2 key to select the white inverted characters “BIN.SUM” (Figure
8-24).
z
Press F3 key to see the test result.
※ Sorting judging rule: To check whether the secondary parameter is within
the setting range, if not, just display SREJ instead of judging the primary
parameter, if yes, continue to proceed primary parameter sorting.
z
Press F4 key to delete test result (Figure 8-25).
z
Press F3 orF4 key to select delete or non-delete test result (Figure8-26).
z
Press F2 key to see next page of test result (Figure 8-27, 8-28).
SORT BY C
NOM_VAL:15.000pF
D_MIN: .0010 D_MAX: 1.000
TEST RESULT
FAIL_ITEM
TOTAL
PHI ( C >MAX )
PLO ( C <MIN )
SREJ( D NG )
2
TOTAL
2
PASS_ITEM
BIN 1 - BIN 13
EXIT
63
Press F3 to
delete the
testing data
YES
NO
Figure 8-26
62
NEXT
PAGE
Press F4 to
maintain the
testing data
LCR-800 Series User Manual
LCR-800 Series User Manual
9. OPTION 2 (RS-232C: option for LCR-816/817/819, standard
for LCR-821)
SORT BY C
NOM_VAL:15.000pF
D_MIN: .0010 D_MAX: 1.000
TEST RESULT
FAIL_ITEM
TOTAL
PHI ( C >MAX )
PLO ( C <MIN )
SREJ( D NG )
2
TOTAL
2
PASS_ITEM
BIN 1 - BIN 13
NEXT
PAGE
Press F2 to
next page
9-1. On-line Procedure
1)
Power on the LCR METER.
2)
Power on the RS232 function of LCR METER. With the steps as
follows (Figure 1):
z Press MENU key.
CLR.
z Press F2 key to select “SORT” function.
EXIT
z Press F3 key to select “OPTION 2” function.
z Press F1 key to select “RS232” ON.
Figure 8-27
3) Run PC LCR-VIEWER Program.
BIN
1
2
3
4
5
6
7
8
9
10
11
12
13
MIN pF
13.500
12.750
12.000
11.250
10.500
9.750
9.000
8.250
7.500
6.750
6.000
5.250
4.500
MAX pF
16.500
17.250
18.000
18.750
19.500
20.250
21.000
21.750
22.500
23.250
24.000
24.750
25.500
TOTAL
999999
999999
999999
999999
999999
999999
999999
999999
999999
999999
999999
999999
999999
4) Check the contents of the Message area under the LCR-VIEWER to
LAST
PAGE
make sure the Online Procedure has been well done. If not, select the
setting items above the LCR-VIEWER to change the Com port and
make connection again until it is online. Please refer to Figure 9-2, 93, 9-4.
EXIT
Figure 8-28
64
65
LCR-800 Series User Manual
LCR-800 Series User Manual
SPEED
SLOW
C 13.450uF
D .1652
DISPLAY
VALUE
TESTING
F :1.0000 kHz R.H OFF
V :1.000 V
C.V OFF
AUTO
MANU INT.B OFF
CAP. C/D OFFSET
MODE
C/D
CIRCUIT
SERIES
MENU
OFFSET
SET SORT
Press menu
key
SORT
SET PARAMETER
SETTING
CALIBRATION
CALBRAT
Press F2 key
to the sort
menu
EXIT
Figure 9-2
NOM.VAL
77.000
NOM.VAL = 77.000pF
OPTION2
RS232
Press F3 key to
input the
option2
function
EXIT
RS232 INTERFACE:ON
ON
OFF
Press F1 key
to the RS232
ON
EXIT
Figure 9-3
Figure 9-1
66
67
LCR-800 Series User Manual
LCR-800 Series User Manual
9-2. RS232 VIEWER Software Operation
1) File
Press Exit(Figure 9-6), or press Power to leave the program.
Figure 9-4
5) After online, the picture of LCR Meter will be switched to “RS232
ONLINE”, please refer to Figure 9-5.
RS232 ONLINE
Figure 9-6
Figure 9-5
68
69
LCR-800 Series User Manual
LCR-800 Series User Manual
2) Option
Settings
Port:
There are three Ports available for selection including
Com1, Com2 and Com3. The default value is Com1.
Baudrate:
After online, there are five Baudrates available for
selection including 9600, 19200, 38400, 57600 and
115200. The default value is 38400。
DataBits:
There are 8 bits of Data bits and User can not change it.
Parity:
None. User can not change it.
StopBits:
There is only 1 Bit of StopBits, and User can not change it
(Figure 9-7).
3) Set Filename
This filename setting is the route for saving test result.
Driver: Set driver。
Directory:
Set directory。
File_Name:
File_Num:
Set file name with 4 alphabets or 4 numbers input.
Set 4 codes of file numbers from 0001 to 9999. When
the test results data reaches to the number of 10000 can
be stored with a file number.
Test Result File_Name File_Num Filename
1-10000
LCR_
0001 LCR_0001.Txt
10001-20000 LCR_
0002 LCR_0002.Txt
20001-30000 LCR_
0003 LCR_0003.Txt
30001-40000 LCR_
0004 LCR_0004.Txt
40001-50000 LCR_
0005 LCR_0005.Txt
50001-60000 LCR_
0006 LCR_0006.Txt
|
99980001-99990000 LCR_ 9999 LCR_9999.Txt
FileNum Reset (refer to the following figure): File_Num is 0001.
Ps. File_Num will be continued from last file number of test result. If
last file number is end at 0006, next file number is started from 0007
when re-power on the unit.
If want to start from 0001, reset the FileNum (refer to Figure 9-8).
Figure 9-7
Figure 9-8
70
71
LCR-800 Series User Manual
LCR-800 Series User Manual
4) Data
Result : Display test results. When the test results data reaches to the
number of 10000 will be stored in a file automatically. If want
to store the data less than the number of 10000, must exit the
LCR-VIEWER first (the data can be stored automatically),
then execute again the LCR-VIEWER to start another
counting of test result data. Please refer to Figure 9-9 and 910.
Figure 9-10
Figure 9-9
72
73
LCR-800 Series User Manual
9-3. The Configuration of Cable
Use the cable between DCE and DTE.
9 PIN D-SUB FEMALE to Computer (D-SUB1)
9 PIN D-SUB FEMALE to LCR Meter (D-SUB2)
D-SUB 1 D-SUB 2
Receive Data
2
3
Transmit Data
Transmit Data
3
2
Receive Data
Data Terminal Ready
4
6+1
System Ground
5
5
System Ground
6+1
4
Data Terminal Ready
Request to Send
7
8
Clear to Send
Clear to Send
8
7
Request to Send
Data Set Ready + Carrier
Detect
74
Data Set Ready + Carrier
Detect