Download User`s Manual
Transcript
6.7 Remote Control Examples The following are some examples of ASCII character command sequences: Example 1 B F1 3.141Z N 2.3Z F0 NOTE: You do not need spaces between the characters. They were added here only to make the commands more readable. This command sequence breaks down as follows: B - Set 4071 to Sinewave mode F1 - Move cursor to field 1 (frequency field) 3.141Z - Enter a freq. value of 3.141 MHz N - Move cursor to next cursor field (field 2, level field) 2.3Z - Enter a level of +2.3 dBm F0 - Move cursor to field 0 (turn cursor off) Example 2 M6 F1 0 F2 432X F3 5z f4 10y f0 O -1.23Y T Note: All commands are case insensitive; you may use either upper case or lower case letters. This command sequence breaks down as follows M6 - Set 4071 to Burst mode F1 - Move cursor to field 1 (Trigger/Continuous field) 0 - Set Triggered burst mode F2 - Move cursor to field 2 (on time/duration field) 432X - Enter a duration value of 432 mS F3 - Move cursor to field 3 (burst frequency field) 5z - Enter a freq. value of 5 MHz f4 - Move cursor to field 4 (off time/delay field) 10y - Enter a delay value of 10 seconds f0 - Move cursor to field 0 (turn cursor off) O - Go to offset entry -1.23Y - Enter an offset value of -1.23 volts T - Trigger the burst Example 3 A F1 LLLL UUU f0 This command sequence breaks down as follows A - Reset 4071 to Sinewave mode F1 - Move cursor to field 1 (frequency field) LLLL - Move the cursor left 4 times (to the 1,000 Hz digit pos.) UUU - Increment this digit 3 times f0 - Move cursor to field 0 (turn cursor off) 45 BK Precision 4071 User Manual Rev.2.2