Enhanced Maintenance Routine Enhanced Maintenance Routine Description of Routine The Enhanced Maintenance routine clears any bubbles from the system and clears any clogs from the probe and fluidics. When to Run this Routine This routine is recommended when a low bead count occurs during a protocol run or when calibration or verification fails. Estimated Run Time 14 minutes Before Running the Routine Before you run the Enhanced Maintenance Routine, Bio-Rad recommends that you check the following: 1. Check that the correct calibration and control lots are being used. 2. Clean the sample probe. 3. Check that the waste container is not full. 4. Run the Adjust Probe Height procedure. To Run the Enhanced Maintenance Routine 1. Navigate to the Maintenance view. 2. In the Select Routine section, select Enhanced Maintenance. 3. Add 5 ml distilled or deionized water, 5 ml 70% isopropanol, and 5 ml 0.1N NaOH to the appropriate reservoirs in the reagent block. 4. Click Start. User Guide | 79