Download 7890 Series GC Troubleshooting
Transcript
3 Chromatographic Symptoms Peaks Not Displayed/No Peaks • If using an autosampler: • Ensure that there is sample in the vial. • Verify that the ALS plunger carriage is secured to the syringe plunger. • Check that the syringe is installed correctly and draws sample. • Verify that the turret/tray is loaded correctly and injections are not from out- of- sequence vials. • Watch to see that the sample is pulled into the syringe. • Verify the detector in use is assigned to a signal. • Check the column for proper installation. • Ensure that the column is not plugged. (See “To Measure a Column Flow”.) Perform column maintenance. • Check for leaks. (See “Checking for Leaks”.) • Check the flow settings, then measure the actual detector flows. (See “To Measure a Detector Flow”.) • Some samples change with heat or ultraviolet light. Check sample stability. • Check the sample level in the vial. • If the sample is viscous, try the following: • Increase the viscosity delay time. • Dilute the sample in an appropriate low- viscosity solvent. • Turn the tower fan off. • For 7683B ALS, use the tray quadrants to warm the sample. For 7693A ALS, use the vial heater (accessory G4514A Bar Code Reader/Mixer/Heater) to warm the sample vial. If the problem is with the detector, see Table 3. 48 Agilent 7890 Series Troubleshooting