Chapter 4.4
MVV Measurement
4.4.3 MVV Measurement Guidelines
The measurement should use "running start" to ensure best results
In order to inhale the absolute maximum volume, the breathing frequency should not be too fast
According to known clinical practice, the optimum is about 30 cycles per minute
It is important that the patient uses a nose clip, forms a tight seal around the mouthpiece with his or
her lips and breathes maximally throughout the recording phase
4.4.4 End Of Measurement And Evaluation
Selection of the
computed results.
Icon or pressing the [Enter] key ends the measurement and displays the
4.4.5 MVV Window After Measurement
Selecting the
Icon or pressing the [F6] key toggles the results window on and off.
The content of the results window can be configured by double clicking on the results window. This
leads into the parameter selection window, where parameters can be selected to be displayed in the
results window for future tests. The selection can be changed at any time.
ZAN500 Body
73 - 203
Part No. 5001005ENG.A