Download Thermo Scientific AquaSensors
Transcript
Calibration and Configuration | Section 7 From the Main Menu, use the up/down arrows to select Calibrate. Then press ENTER. ≡Main Menu ►Calibrate 1 ▲ MENU ESC ENTER ◄ ► ▼ From the Calibrate Menu, use the up/down arrows to select 1-Point Sample. Then press ENTER. ≡Calibrate ►1-Point Sample ▲ MENU ESC ENTER ◄ ► ▼ The current sensor value is dynamically displayed and asks the user to be sure the measurement is ready for calibration before proceeding. When ready, press the ENTER key. 8.17 ppm Sensor ready? ▲ MENU ESC ENTER ◄ ► ▼ The calibration value can be adjusted with the arrow keys. Use the up/down arrows for fine adjust and the left/right arrows for coarse adjust. After the calibration value is adjusted to the desired value, press the ENTER key. 1-Point Sample? ( 8.17) ppm ▲ MENU ESC ENTER ◄ ► ▼ Thermo Scientific AquaSensors RDO Pro-X Optical Dissolved Oxygen Measurement System | 33