Download Instruction manual
Transcript
MI3002 MI3100 MI3102 Instrument operation Fuse F1 (continuity circuit) blown or not inserted. SF Single fault condition in IT system. Noise voltage is present between H and E or S test terminals. Resistance of auxiliary earth electrode is higher than 100×RE. Check the auxiliary earth electrode. Probe resistance is higher than 100×RE. Check the probe. Resistances of auxiliary earth electrode and probe are higher than 100×RE. Check all probes. 4.1.4 Result field Measurement passed. Measurement failed. Measurement is aborted. Check the conditions at the input terminal. 4.1.5 Other messages Hard Reset No probe Instrument settings and measuremet parameters/limits are set to initial (factory) values. For more information refer to chapter 4.5.5. Recalling original settings. LUXmeter probe is turned off or disconnected from the EurotestXE instrument. Connect probe to the instrument using RS232 connector and turn it on. First measurement First stored measurement results are displayed. Last measurement Last stored measurement results are displayed. Memory full Already saved CHECK SUM ERROR All memory locations are occupied. Measurement results already saved. Memory contents damaged. Contact your distributor or manufacturer for further information. 21