Chapter 7 | Troubleshooting
Cont.
94 | Varioskan LUX
Code
Explanation
Suggested action
80
Measurement chamber door is open.
See that nothing is obstructing the
measurement chamber door from closing
when the plate is driven in. Do not open the
door while the instrument is measuring.
81
Any error during start-up preventing the execution of
some commands. 1
The measurement chamber door must be
closed during start-up. If it is not, contact
service.
82
The number of filter pairs does not match the number of
delay/integration time pairs. This error may be reported
in response to a measurement command when a TRF
measurement is selected.
Make sure that the number of filter pairs in a
FIL command and the number of delay/
integration time pairs in a SET 8 command are
equal.
83
LAT module aperture plate position is incorrect.
Contact service.
84
LAT module reference LED failure.
Contact service.
85
Dispensing tip is in wrong position.
When priming or washing a dispenser, the
dispensing tip must not be inserted into any
dispensing position. When dispensing or
carrying out tip priming, the dispensing tip
must be inserted into one of the dispensing
positions.
86
The command cannot be executed for the wrong plate
type.
Use a suitable plate type. DME and DIS
commands only work with plate types having
up to 384 wells.
87
A dispenser has run out of liquid.
Add liquid and prime the dispenser.
88
LAT module base is installed but the LAT module itself is
not.
Install the LAT module or replace the base
with a dual dispensing base or a dummy.
89
LAT module position calibration failed.
90
LAT module position calibration failed.
91
TRF lamp fail.
LAT module flash lamp failure.
92
Varioskan LUX LED fail.
LAT module Varioskan LUX LED failure.
93
Integrated gas module is not installed.
Gas control feature not available.
94
Integrated gas module communication error.
Errors detected in communication with the
Integrated gas module.
95
Gas concentrations too high.
96
Integrated gas module memory error.
Integrated gas module memory access failed
or data was corrupted.
97
CO2 concentration above target.
Target CO2 concentration above target.
98
CO2 concentration below target.
Target CO2 concentration below target.
99
CO2 sensor failure.
CO2 sensor failure.