Download Varioskan LUX Technical Manual
Transcript
Chapter 7 | Troubleshooting Cont. 94 | Varioskan LUX Code Explanation Suggested action 80 Measurement chamber door is open. See that nothing is obstructing the measurement chamber door from closing when the plate is driven in. Do not open the door while the instrument is measuring. 81 Any error during start-up preventing the execution of some commands. 1 The measurement chamber door must be closed during start-up. If it is not, contact service. 82 The number of filter pairs does not match the number of delay/integration time pairs. This error may be reported in response to a measurement command when a TRF measurement is selected. Make sure that the number of filter pairs in a FIL command and the number of delay/ integration time pairs in a SET 8 command are equal. 83 LAT module aperture plate position is incorrect. Contact service. 84 LAT module reference LED failure. Contact service. 85 Dispensing tip is in wrong position. When priming or washing a dispenser, the dispensing tip must not be inserted into any dispensing position. When dispensing or carrying out tip priming, the dispensing tip must be inserted into one of the dispensing positions. 86 The command cannot be executed for the wrong plate type. Use a suitable plate type. DME and DIS commands only work with plate types having up to 384 wells. 87 A dispenser has run out of liquid. Add liquid and prime the dispenser. 88 LAT module base is installed but the LAT module itself is not. Install the LAT module or replace the base with a dual dispensing base or a dummy. 89 LAT module position calibration failed. 90 LAT module position calibration failed. 91 TRF lamp fail. LAT module flash lamp failure. 92 Varioskan LUX LED fail. LAT module Varioskan LUX LED failure. 93 Integrated gas module is not installed. Gas control feature not available. 94 Integrated gas module communication error. Errors detected in communication with the Integrated gas module. 95 Gas concentrations too high. 96 Integrated gas module memory error. Integrated gas module memory access failed or data was corrupted. 97 CO2 concentration above target. Target CO2 concentration above target. 98 CO2 concentration below target. Target CO2 concentration below target. 99 CO2 sensor failure. CO2 sensor failure. Thermo Scientific