The devices can now be turned on in the following order: • First turn on vacuum pump • After, turn on the PoleStar cold trap (see picture 1.11) • Finally turn on the CombiDancer (see picture 1.3). When using the Vacuubrand PC-2003/2004 VARIO Diaphragm Pump, the additional steps mentioned in section 1.5.2 must be carried out, in order to allow remote operation of the pump, i.e. the pump can be controlled directly through the CombiDancer. The system is then ready to operate. 1.5.2 Activation of remote operation for the Vacuubrand PC-2003/2004 VARIO Diaphragm Pump To control and operate the Vacuubrand PC-2003/2004 VARIO diaphragm pump directly through the CombiDancer, following steps must be carried out once on the pump: • Keep the second arrow button V from the left pressed while turning the device on. Controller opens the setup program for interface setup Display: Set b.96 • Press Start button Display: Set d.8 • Press Start button Display: Set P.no • Press Start button Display: Set h.no • Press Start button Display: Set r.no • Press the second arrow button from the left to activate remote operation Display: Set r.On • Press Start button • Controller is initialized (Display: Software Version) • Pressure in mbar and PC are displayed CAUTION: Always turn pump and cold trap on before the CombiDancer; otherwise no communication between the devices will be possible. (Error 30) 14