Digi-Pas TM
A Ventura Group of Companies
Applications
Digi-Pas TM Digital Level
ON/OFF: Turn on/off device
USER MANUAL
LCD Digital Display
DIGITAL LEVEL
DWL80E (0.1° Resolution)
® Registered Design
Product of Singapore
CALIBRATION INSTRUCTIONS
Aluminium
Structure
Alternate Zero
Magnet
MODE: Select 0.1° or %
Freeze Display
value & store the
value in the memory
® Registered Design
For first time usage, remove the plastic strip to
connect the battery.
CALIBRATION PICTURE
STEP 1
1.(a). Push Once on the CAL button, “CAL 1” blinks.
Vial
CAL: Calibrate the device
calibration value
The Digi-Pas™ Digital level is warranted to the original
purchaser to be free from defects in workmanship and material.
JSB Tech will, at its option, repair or replace any detective part
which may malfunction under normal and proper use within a
period of 1 (one) year from the date of shipment. The foregoing
warranty shall not apply to defects resulting from misuse, abuse
or transfer by the Buyer, Buyer-supplied software or interfacing,
unauthorized modification or operation outside of environmental
specifications for the product. JSB Tech does not warrant that the
operation of the instrument software, or firmware, will be
uninterrupted or error free. The exclusive remedy under any
and all warrants and guarantees, expressed herein, and we shall
not be liable for damages from loss or delay of equipments uses,
consequential, or incidental damage. No other warranty is
expressed or implied. JSB Tech specifically disclaims the
implied warranties of merchantability and fitness for a
particular purpose.
(b). Place DWL80E on a flat & clean surface
(e.g. Table)
2.(a). Press MODE, display counts down from
10,9,8.......1, “CAL 2” blinks.
E.g Table Surface
Push Once “ CAL”
STEP 2
(b). Place DWL80E reverse horizontal (180°) Press
MODE, display counts down, “CAL 3” blinks.
a.
CAL
(c). Place DWL80E vertically (-90°). Press MODE
button, display counts down, “CAL 4” blinks.
(d).Place DWL80E reverse vertically (+90°). Press
MODE button, display counts down.
WARRANTY
Press Mode button Once
b.
180°
CALIBRATION DONE.
NOTE: Ensure DigiPas device is kept still during
calibration.
CAUTION
c.
90°
Manufactured by
JSB TECH Pte Ltd
90°
Press Mode button Once
Press Mode button Once
CAL
DigiP as device is calibrated in the factory for
maximun accuracy. User can re-calibrate when
necessary.
Remove Screw to remove battery from the