Download Instruction Manual
Transcript
Zero-set OFF This function cancels the zero set for measurement values. Setting Function Reset The zero set is not canceled. Hold The zero set is canceled. Panel display • External multi input (MI) can turn the zero set ON and OFF. ( Functions and Arrangements of I/O Terminal Block.") refer to "2-1 ● Setting Procedure 1 Display the "Data Processing Setting" screen while the system is in setting mode. 2 Press the [ENTER] key. The "Average Time" screen will be displayed. 3 Press the [DOWN] key four times. The "Zero-set OFF" screen will be displayed. 4 Press the [ENTER] key. "Yes/No" will be displayed. 5 Press the [UP]/[DOWN] key to make "Yes/No" changes. (Example: Maintaining the zero-set state.) 6 Press the [ENTER] key. Phone: 800.894.0412 - Fax: 888.723.4773 - Web: www.clrwtr.com - Email: [email protected]