Download Untitled - BeneCheck
Transcript
The screen will display the control test results in Hb(g/dL or mmol/L) and Hct(%) alternately. Compare the results to the control range listed on the test strip vial. 6. Hint : Hct is calculated by multiplying 3.00 to Hb Hct % Note : ● Do not reuse test strips. ● Newly opened vials of control solution and test strips must be marked with an opening date. ● Close the test strip vial lid tightly after use. ● Do not use BeneCheck Hb Test System to do any test until you can get the control test result to fall within the range listed on the test strip vial. ● If the test results fall outside the control range, please refer to Chapter 8 for trouble shooting. 19