Site of installation for the transmitter
When determining the site of installation, it is important to know the exact ambient conditions and
to take them into consideration. In order to receive representative results, the ventilation conditions
must be taken into account.
The position of the transmitter in the room must ensure that the gases still reach the sensor even in
the event of unfavorable ventilation. If necessary, carry out a measurement, e.g. with ventilation
smoke tubes.
When specifying the site of installation, it must be further observed that the transmitter can be
accessed for service and calibration tasks. The transmitter must be installed in a horizontal position,
with the measuring chamber on the left or right.
External influences must also be considered, such as:
• rainwater, splash water, dripping water, condensation
• the dust content in the atmosphere
The transmitter is protected against the ingress of water or dust to the greatest possible extent.
Special accessories can protect the transmitter against damage in extremely harsh measurement
conditions. If desired, GfG will gladly advise you about suitable measures.
!
The warranty may become void if the sensor is exposed to ambient conditions which
were unknown to GfG Instrumentation during the planning phase or delivery.
Assembly
Assembly takes place step-by-step in the following sequence:
5