Download AFG1000 Series Arbitrary Function Generator Quick Start User Manual
Transcript
Operating basics 3. 4. 5. 6. 7. 8. Press Sweep Time to set the time between the start and stop frequency. Press Linear/Log to select the sweep type. Press StartFreq/ CtrFreq. Use the general purpose knob or the numeric keypad to set the start or the center frequency. Press StopFreq/ FreqSpan. Use the general purpose knob or the numeric keypad to set the stop frequency and the frequency span. Press NextPage to enter next submenu. Press Trigger to choose a source between internal, external, and manual. External defines the source input from the Fsk/Ext Trig In interface. Press Slope to switch between Positive and Negative. Manual defines starting the sweeping whenever the general purpose knob is pushed. NOTE: All the parameters can be adjusted by the general purpose knob or the keypad. Sweep frequency facts If a start frequency is lower than a stop frequency, the instrument sweeps from the low frequency to the high frequency. If a start frequency is higher than a stop frequency, the instrument sweeps from the high frequency to the low frequency. Once the sweep is selected, the frequency is swept from the sweep start to the sweep stop frequencies. AFG1000 Series Quick Start User Manual 35