Download Manual - Fondriest Environmental
Transcript
ORP voltage TruLab pH/ISE 1320 Measured parameter Time interval Stability in the time interval ORP 15 seconds ∆ : better than 0.3 mV Temperature 15 seconds ∆ : better than 0.5 °C The minimum duration until a measured value is assessed as stable is the monitored time interval. The actual duration is mostly longer. 7.1.3 Measuring the temperature For reproducible ORP measurements, it is essential to measure the temperature of the test sample. You have the following options to measure the temperature: Automatic measurement of the temperature with the temperature sensor (NTC 30 or Pt 1000) integrated in electrode. Measurement by an external temperature sensor. Manual determination and input of the temperature. The measuring instrument recognizes whether a suitable sensor is connected and automatically switches on the temperature measurement. The display of the temperature indicates the active temperature measuring mode: Temperature sensor Resolution of the temp. display Temp. measurement yes 0.1 °C Automatic with temperature sensor - 1 °C Manual If you wish to measure (or calibrate) without temperature sensor, proceed as follows: 1. Measure the current temperature of the test sample. 2. Set the temperature value with <><>. or In the <MENU/ENTER>/U/Man. temperature menu, set the temperature value with <><>. 7.2 ORP calibration ORP electrodes are not calibrated. You can, however, check ORP electrodes by measuring the ORP of a test solution and comparing the value with the nominal value. 44 ba76157e01 03/2014