Download - Frank`s Hospital Workshop

Transcript
Sonicaid FM800 Series Reference Manual
Maternal blood pressure
Method
Oscillometric
Measurement ranges
Systolic 50–280 mmHg
Diastolic 10–260 mmHg
Pulse 40–240 bpm
Accuracy
±2% or 3mmHg, whichever is greater
Modes
Manual or automatic
User-selectable interval in Auto Mode:
3, 5, 10, 15, 20, 30, 45, 60, 90 or 120 minutes
Record/display
On-screen display and printed record of:
Systolic blood pressure
● Diastolic blood pressure
● Pulse rate
Printed record of:
●
●
Mean arterial pressure
[Note: Mean arterial pressure not shown in the USA or Canada]
Alarms
Systolic High: 55–255 mmHg
Systolic Low: 50–250 mmHg
Diastolic High: 35–220 mmHg
Diastolic Low: 30–215 mmHg
System: air leaks. movement, overpressure etc
Safety
Type CF protection
Hardware and software controls to limit:
●
●
Standards
102
Inflation (max. 300 mmHg)
Measurement time (max. 160 secs)
ANSI/AAMI SP10-1992