Download ViSolid
Transcript
ViSolid® 700 IQ (SW) Indexes 8 Indexes 8.1 Explanation of the messages This chapter contains a list of all the message codes and related message texts that can occur in the log book of the IQ SENSORNET system for the ViSolid® 700 IQ (SW) sensor. Information on the contents and structure of the log book and the structure of the message code is given in the LOG BOOK chapter of the IQ SENSORNET system operating manual. All message codes of the ViSolid® 700 IQ (SW) end with the number "342". 8.1.1 Error messages Message code Message text EA2342 Sensor temperature too high! * Check process and application EA3342 Sensor temperature too low! * Check process and application EA6342 Meas. range exceeded or undercut * Check process * Select other meas. range * Submerse sensor in sample * Select bubble-free spot for measurement * Remove any foreign matter from sensor * Avoid influence of large foreign matter * Clean sensor * Increase signal average time EA7342 Ultrasound cleaning system switched off * Check sample temperature * Submerse sensor in sample ba76040e04 05/2014 8-1