Download Leksell® Multi Purpose Stereotactic Arc
Transcript
1 L E K S E L L® M U L T I P U R P O S E S T E R E O T A C T I C A R C 2 3 4 Quick reference guide Leksell® Multi Purpose Stereotactic Arc 4 3 40 30 2 20 10 0 11 -10 90 -20 -40 1 6 8 5 FT 9 0 LE 1 2 0 90 60 150 100 RIGHT 90 00 10 1 180 X 10 20 1 30 10 80 POST 60 50 40 210 Y 16 140 150 A NT 9 0 R IG H T 24 0 300 270 110 100 10 90 80 9 Leksell® Multi Purpose Stereotactic Arc is part of Leksell Stereotactic System®, which is intended for localization and diagnosis of intracranial disorders and their surgical treatment including radiosurgery and stereotactic radiation therapy. 330 7 Introduction This is a quick reference guide for Leksell® Multi Purpose Stereotactic Arc. It is the user’s responsibility to be fully informed of its proper use. For detailed instructions read Leksell® Multi Purpose Stereotactic Arc, Instructions for Use. 70 60 50 40 Z Caution The user must read and fully understand the Hazard Notices and other Notices contained in Leksell® Multi Purpose Stereotactic Arc, Instructions for Use. Names and abbreviations Arc = Leksell® Multi Purpose Stereotactic Arc Main Parts 1 Semicircular arc 2 Instrument carrier 3 Instrument stop holder – engraved 6 Axis support 7 Arc axis – engraved 8 Arc axis 9 Arc supports – engraved 4 Instrument guide holder 10 Slides 5 Axis support – engraved 11 Instrument screwdriver Frame = Leksell® Coordinate Frame G 1 2 3 4 L E K S E L L® M U L T I P U R P O S E S T E R E O T A C T I C A R C Arc orientations Assembling – lateral arc orientations Lateral right orientation • The engraved arc axis is on the patient’s right side. Fitting arc supports and setting Z coordinate 1. Insert an arc support 9 in a slide 10 • The engraved surface of the arc faces anterior. 2. Move support and align the required Z coordinate exactly with the index mark 12 • The X coordinate is set on the RIGHT scale. 120 110 12 100 90 10 80 70 60 ��� ��� ��� �� �� �� LEFT �� ��� ��� �� �� ��� �� �� 20 110 100 9 �� ��� 160 150 140 ANT RIGHT 10 100 90 80 POST X 60 50 40 Y X R R L �� �� �� �� �� ��� ��� ��� ��� ��� ��� Lateral left orientation • The engraved arc axis is on the patient’s left side. • The engraved surface of the arc faces posterior. • The X coordinate is set on the LEFT scale. 3. Secure the arc support at the required Z coordinate. 9 NOTE Do not overtighten the screw to secure Z coordinate. Otherwise it may be difficult to move slide and arc support along the frame. ��� ��� ��� ��� �� �� �� �� �� �� � L ��� ��� ��� �� ��� ��� �� �� �� �� �� LEFT �� ��� �� �� RIGHT 20 110 100 9 ��� 160 150 140 ANT 4. Fit and secure the other arc support in the slide in the same manner. 10 100 90 80 POST X L 60 50 40 Y X R ��� ��� ��� ��� ��� ��� ��� �� �� �� �� �� �� Sagittal posterior orientation • The engraved arc axis is at the patient’s posterior. • The engraved surface of the arc faces the patient’s right side. Fitting slides and setting Y coordinate 1. Insert a slide (with arc support) 10 over the �� ��� � ��� groove of a side bar. ��� ��� ��� �� 10 �� �� • The Y coordinate is set on the POST scale. �� �� ��� ��� �� �� ��� ��� ��� ���� ��� � � �� �� �� �� � P ���� ����� ��� ��� ��� ��� � A P Y � �� �� �� �� �� �� �� ��� ��� ��� ��� ��� ��� ��� ��� 2. Move the slide and align the required Y coordinate exactly with the vertical index mark 13 ��� ��� ��� �� �� � ��� �� �� �� �� Sagittal anterior orientation • The engraved arc axis is at the patient’s anterior. 13 �� • The engraved surface of the arc faces the patient’s left side. • The Y coordinate is set on the ANT scale. 3. Secure the slide with the instrument screwdriver. A 130 120 110 30 40 50 100 90 80 Y X ANT 120 110 100 POST Y LEFT RIGHT 160 150 140 130 1 A � ������������� ��� ��� ��� ��� ��� ��� ��� ��� �� �� �� �� �� �� �� P ��� � 4. Fit and secure the other slide with arc support. 100 90 80 70 100 1 120 130 Y 140 150 160 170 0 70 60 50 40 1 L E K S E L L® M U L T I P U R P O S E S T E R E O T A C T I C A R C Fitting arc axis and setting X coordinate Ensure the small tab fits into the corresponding indent. 15 4. Carefully push the unmarked arc axis inwards and hold it firmly. Tighten thumbwheel on back of unmarked arc 16 axis with other hand to secure the unmarked arc axis in place. Make sure there is no narrow opening 16 20 100 90 LEFT 20 110 G-frame 10 30 0 60 5 Y 50 RIGHT POST 90 80 10 100 X 40 Y 30 90 0 120 3 0 0 0 RIG HT 150 270 240 210 1 8 0 60 50 40 30 25 20 40 ONLY FOR G-frame X 30 0 60 5 30 50 40 90 0 120 3 0 0 270 0 RIG HT 150 160 90 RIGHT 10 100 X 80 POST Y 0 RIGHT 0 60 50 40 • Lateral right orientation – the engraved arc axis must mate with the arc support ring on the patient’s right side. • Lateral left orientation – the engraved arc axis must mate with the arc support ring on the patient’s left side. 20 40 160 10 90 RIGHT 10 100 X 5 0 RIGHT Y ST 60 80 PO 50 40 Ensure the small tab fits into the corresponding indent. ONLY FOR G-frame 240 1. Align the arc correctly with the frame. 210 1 8 Y 140 ANT 160 150 10 5 100 90 LEFT 20 110 O S T 60 80 P 100 90 R IG HT 10 ONL 7. Tighten the thumbwheels on both arc axis to secure the arc at the desired ring angle. 160 20 FT 100 90 LE 20 110 ame Y FOR G-fr Fitting the arc and setting the ring angle FT 100 90 LE 20 110 25 25 40 140 A NT 160 150 140 ANT 160 150 ONLY FOR 40 140 ANT 160 150 30 14 2. Introduce the ring of the engraved arc axis into the annulus of the ring on the arc support. 5. Release the thumbwheel at the ring on each arc axis. 6. Tilt the arc until the index mark is exactly aligned with the desired ring angle. NOTE Make sure to use correct scale. 3. Secure the arc axis by tightening the screw 14 on the engraved face with the instrument screwdriver. LEFT 30 100 9 20 110 ANT X 1. Loosen thumbwheel at back of unmarked axis support and slide unmarked axis to open position. O S T 60 80 P 100 90 R IG HT 10 15 2. Move the engraved arc axis and align the required 160 150 140 X coordinate exactly with the X index mark. 3 3. Fit the ring of the unmarked arc axis into the ring of the other arc support. 160 NOTE If the arc axis holders have been disassembled ensure that both have been properly remounted. There should be no gap 15 between them and the back surface of the arc. 2 NOTE It is very important that the two thumbwheels are properly tightened. Assembling – sagittal arc orientations NOTE For sagittal orientations use a straight or slotted front piece – not a curved one. Assembling of the arc for sagittal orientations follows in principal the same workflow as for lateral orientations with some important exceptions. 1. Setting Z coordinate The Z coordinate is set in the same way as for lateral orientations. 130 120 The coordinate is set on the arc support. 110 100 90 80 70 60 50 40 Z 4 4 L E K S E L L® M U L T I P U R P O S E S T E R E O T A C T I C A R C ��� ��� ��� ��� ��� ��� ��� �� � ��� �� � � �� ��� �� � �� �� � ��� �� ��� �� �� � �� 3. Setting Y coordinate For sagittal orientations the Y coordinate is the last coordinate set. 1. Insert the stem of the instrument guide holder 19 into the unmarked slot in the instrument carrier with the collar nearest to the patient’s skull. 2. Tighten the thumbwheel to secure the guide holder. 7 20 40 G-frame X Y 140 ANT 160 150 ONLY FOR FT 100 90 LE 20 110 3. Insert the stem of the instrument stop holder 20 in the marked slot of the instrument holder with the collar nearest to the patient’s skull. 5 0 RIGHT 50 40 PO ST 60 The coordinate is set on the engraved arc axis 7 with the Y index mark. Fitting the instrument guide and stop holders NOTE Make sure to use the correct scale. 4. Align the arc correctly with the frame. • Sagittal posterior orientation – the engraved arc axis must mate with the arc support ring on the patient’s posterior. • Sagittal anterior orientation – the engraved arc axis must mate with the arc support ring on the patient’s anterior. Fitting the instrument carrier Setting the arc angle 4. Move instrument stop holder until the horizontal index mark on the carrier is exactly aligned with the desired mark on the stop holder scale. 5. Tighten the thumbwheel to secure the stop holder in place. 20 19 6. The arc is now fully assembled. The stop insert 21 and guide insert 22 of the selected instrument can now be inserted into the collars. See User’s manual for selected instrument! -40 �� 21 22 Removing of the Arc 1. Release locking thumbwheel 18 Remove the Arc from the Frame in the reverse of that when securing the Arc. 2. Attach the instrument carrier to the arc. Disassembly 3. Ensure that the carrier is correctly oriented so that index mark is visible at the engraved face of the arc. Disassemble the Arc in the reverse of the assembly procedure. 4. Move the carrier until the desired arc angle is exactly aligned with the index mark. For detailed instructions read Leksell® Multi Purpose Stereotactic Arc Instructions for Use. Care & maintenance Elekta Instrument AB. Article no. 014612.00 DEC03. Printed in Sweden. © All rights reserved. No document may be reproduced in any form without written permission from the copyright holder. The information in this document is subject to change without notice and does not represent a commitment on the part of Elekta. The Elekta logotype, Leksell® and Leksell Stereotactic System® are registered trademarks of Elekta AB (publ) or it subsidaries ��� �� -30 �� 0 �� -20 �� 10 �� 18 -10 �� 5. Tighten thumbwheel on top to secure the instrument carrier. 20 2. Setting X coordinate For sagittal orientations the X coordinate is set before setting the Y coordinate. The X coordinate is set on the front piece and back piece of the frame. 30 3 40 2 90 80 RIGHT 10 100 1