When the calibration status becomes Stable from Unstable, to move the cursor to Save by
pressing Up and Down keypad, then press ENT to save the Span calibration.
Calibration for other gases
A. The calibration of CO,CH4,H2,CnHm and O2 is the same as CO2
B. Standard gas for zero calibration should be N2. Standard gas for span calibration is the same as
corresponding gases.
Part V. Diagnoses
When the cursor moves on the “Diagnoses”, press “ENT” to enter it. You will find Sensor Signal inside.
Set up
1.Parameter
2. History
3. Zero
4. Cali.
5. Diagnoses
6. Coefficient
↑ ↓ Select
ESC
ENT
Sensor signals
Ch
00
01
02
03
04
05
06
07
08
ADC
0000
0000
3025
2192
3809
0000
3038
0000
3898
n
H2
Bat
O2
Ch
09
10
11
12
13
14
15
ADC
3720
3647
3549
3549
3549
3549
3549
n
COtst
CO2ref
CO2tst
CH4
CH4tst
CnHmref
CnHm
Press
COref
↑ ↓ Select
ESC
ENT
Part VI. Coefficient
Under set up interface, to move cursor to coefficient, then enter into following interface