Download SCXI-1120/D User Manual
Transcript
Chapter 3 Signal Connections The input signal range of an SCXI-1120 input channel is ±5 V/ Gtotal referenced to its negative input, where Gtotal is equal to the product of the first-stage and second-stage gains. In addition, the input channels are overvoltage protected to 250 Vrms with power on or off at a maximum of 4.5 mArms sink or source. ♦ SCXI-1120D When connecting AC-coupled signals to the SCXI-1120D, you do not need to add the external resistor, as the SCXI-1120D provides a DC path internally. Refer to Figures 3-6 and 3-7 for further details. 511 kΩ + + + Vs - I + Vcm Vout 511 kΩ - High - CMV Module Figure 3-6. AC-Coupled Signal Connection for the SCXI-1120D with High Common Mode Voltage 511 kΩ + + Vs 511 kΩ - + Vout - I Figure 3-7. Floating AC-Coupled Signal Connection for the SCXI-1120D Temperature Sensor Connection Pins C2 and C4 are dedicated for connecting to a terminal block temperature sensor. The temperature sensor is not isolated and is referenced to the chassis ground. The connection is overvoltage protected to ±25 VDC with power on and ±15 VDC with power off. © National Instruments Corporation 3-7 SCXI-1120/D User Manual