7.2 Application "Percent Weighing"
The "Percent Weighing" application allows you to check a sample weight as percentage to a
reference target weight.
Requirement: The function "PERCENT" must be assigned to the «F» key (see advanced
menu topic "ASSIGN:F".
– Activate function percent weighing "PERCENT" by pressing and
holding the «F» key.
PERCENT
Percent Weighing first requires the setting of a reference weight
that should corresponds to 100%, there are 2 possibilities:
A
A
B
SET 100%
A
Setting the reference in manual mode (enter 100%).
B
Setting the reference in weighing mode (weigh 100%).
Setting possibility
Setting the reference by manual mode (enter 100%)
A
1 Press «
* 175.28
g
» to activate manual mode.
2 Select the reference target weight (100%) by scrolling up («+»
key) or down («–» key). Speed up by press and hold.
3 Press «
» to confirm.
0.0%
B
SET 100%
0 .00
g
Setting possibility
Setting the reference in weighing mode (weigh 100%)
B
1 Press «
» to zero/tare the balance and to activate the
weighing mode. If needed: place empty container on the weighing
pan and zero/tare again.
C
175.28
g
I00.0%
2 Load the reference weight (100%).
Note: Reference weight must be at least +/- 10d.
3 Press «
» to confirm.
Switching between manual mode and weighing mode
,
175. 28
g
– Press «
175. 28
g
Note: By switching from weighing mode to manual mode the weight
value will be transferred and can be changed manually.
» to switch between manual and weighing mode.
Note: If without any key press within 60 seconds, the balance returns to the previous active application.
On completion of the weighing-in procedure, your balance is ready for percent weighing.
40