Download Model 32 / 32B User`s Manual - Cryogenic Control Systems, Inc.

Transcript
Cryo-con Model 32
OVERTEMP commands
These commands are associated with the heater’s Over Temperature Disconnect
feature.
This feature is used to disconnect the heater if a specified temperature is exceeded
on a selected input channel.
OVERTEMP:ENABLE: OTD Enable.
Sets and queries the over temperature disconnect enable.
Command Syntax: OVERTEMP:ENABLE <enab>
Where <enab> is the desired enable status, which may be ON or OFF.
Query Syntax: OVERTEMP:ENABLE?
Query Response: <enab>
Where <enab> is the status of the over temperature disconnect enable.
Command Example: OVERTEMP:ENABLE OFF
Sets the over temperature disconnect feature to OFF.
Query Example: OVERTEMP:ENABLE?
Example Response: YES
Indicating that the over temperature disconnect feature is enabled
Short Form: OVER:ENAB
198