Chapter 5 Remote Interface Reference System-Level Control Commands DIAGnostic :DISPlay[:INFOrmation] <message> :DISPlay:STATe <0|1|OFF|ON> :DISPlay:STATe? These commands control the front-panel display. The DIAGnostic:DISPlay:INFOrmation command is used to write messages to the display. The message parameter is a string that can contain up to 13 characters. You can enter the string in lower case characters, but the display only shows upper case characters. In addition, each character (except the 13th) can be trailed by any one of four special characters (, . : ;). These “trailer” special character are not counted as one of the thirteen characters. Valid characters include:. 5 Type Supported Characters Numeric 0-9 Alphabetic A-Z Symbolic “ ‘ (space) ( ) * + - , . : ; / \ The DIAGnostic:DISPlay:STATe command turns the front-panel display on or off. The ADRS, RMT, and ERROR annunciators may turn on even when the front-panel display is turned off. After an instrument power-on or a reset command (*RST), the display will be turned on. The DIAGnostic:DISPlay:STATe? query returns the display status. The returned value is either a “0” or a “1”, indicating the display is turned off or on, respectively. 158